Keysight NFA Series Performance Verification And Calibration Manual page 48

Noise figure analyzers
Hide thumbs Also See for NFA Series:
Table of Contents

Advertisement

Figure 2-4
Adapter 1
(if Applicable)
Step 9. Select the relevant calibration kit in use from the Cal menu, as the
NOTE
Step 10. Select an S11 port calibration (press the
Step 11. Connect the Open to the test port or, if fitted, the adapter and perform
Step 12. Connect the Short to the test port or, if fitted, the adapter and perform
Step 13. Connect the Load to the test port or, if fitted, the adapter and perform
Step 14. Save the calibration (press the
Test Descriptions
Input VSWR
Input VSWR Test Setup 1
VECTOR NETWORK
ANALYZER 1
Port 1
calibration kit selection is dependant on the NFA being tested (press the
key,
Cal
Calibration Kit
The connector type applies to the test port or adaptor if fitted and not the
NFA under test.
).
Reflection 1-port
the open calibration (press
key).
the open calibration (press
key).
the open calibration (press
).
done 1-port CAL
28
NOISE FIGURE
ANALYZER
Input
Cable 1
selection and
Return
Cal
and then select the relevant menu
Opens
and then select the relevant menu
Shorts
, select the relevant menu key
Load
key and then
Save/recall
Adapter 2
(if Applicable)
).
key, then
Calibrate Menu
Save State
and
and then
).
Chapter 2