NOTE
Table 1-1
Equipment description
Universal Counter
Frequency Standard
Vector Network Analyzer 1
Vector Network Analyzer 2
Type N Calibration Kit
3.5mm Calibration Kit
Preparing for Calibration and Performance Verification Test
Recommended Test Equipment
Recommended Test Equipment
The following tables list the recommended test equipment for the
performance tests. The tables also list recommended equipment for the
Noise Figure Analyzer adjustment procedures.
When performing the performance test manually any recommended
equipment that meets the critical specifications given in the table can be
substituted for the recommended model when manually testing.
Substitution is not permitted when using the automated performance
verification and adjustment software.
Recommended Test Equipment
Critical specification for equipment
substitution
Time Interval Range: 25 ms to 100 ms
Single Operation Range: +2.5 to -2.5 Vdc
Frequency: 10 MHz Timebase Accuracy:
-10
< 1 x 10
10 MHz to 3 GHz
3 GHz to 26.5 GHz
Impedance: 50
UUT: N8972A and N8973A
Impedance: 50
Frequency Band: 10 MHz to 3 GHz
UUT: N8974A and N8975A
6
/day
Ω
Ω
Recommended
model
53132A
5071A
3753ES or
8753ET, opt 004
8722ES or
8722ET, opt 004
85032B
85033D
Chapter 1