Table 2-7
Equipment description
Synthesized Sweeper
(see note)
Adapter 1
Adapter 2
RF Cable
Frequency Standard
BNC Lead (X 2)
NOTE

Related Adjustment

None

Required Test Equipment

Equipment required for Frequency Accuracy test
Critical specification for equipment
substitution
Frequency Range: 10 MHz to 26.5 GHz
Frequency Accuracy (CW): 0.02%
Power Level Range: -55dBm
N8972/3A N Type (M) to 3.5mm (F)
N8974/5A 3.5mm Precision (F) to 3.5mm
(F)
83620/30 3.5mm Precision (F) to 3.5mm (F)
83640/50 2.4mm Precision (F) to 3.5mm (F)
10 MHz to 26.5 GHz
3.5mm (M) to 3.5mm (M), 61 cm (24 in.)
Frequency: 10 MHz Timebase Accuracy
(Aging): < 1 x 10
Length 122cm (48 in.), Frequency 10 MHz
Sythesized Sweeper model 83620B cannot be used test NFA N8975A.
Chapter 2
-10
/day
Test Descriptions
Frequency Accuracy
Recommended model
83620/30/40/50B Option
001 and 008
1250-1744
1250-1749 or 83059B
1250-1749 or 83059B
11901B
11500E
5071A
10503A
39