Table 2-5
NOTE
NOTE
VSWR Frequency Test Bands
PART 5
TEST VSWR
1500 MHz -3000 MHz
PART 6
CAL VNA 3.0 GHz - 6.7
OR 26.5 GHz
PART 7
TEST VSWR
3.0 GHz -6.7 GHz
PART 8
TEST VSWR
6.7 GHz -20.0 GHz
PART 9
TEST VSWR
20.0 GHz -26.5 GHz
•
denotes test required for associated model number.
The NFA uses several frequency dependent paths within the front end
assemblies to cover its operating range. It is therefore not possible to
complete a single VSWR swept measurement at a single fixed frequency
point over its entire operating range, as the NFA automatically switches
at approximately 500 MHz and 3000 MHz, dependant on the fixed
frequency setting.
Chapter 2
N8972A
N8973A
•
Test Descriptions
Input VSWR
N8974A
N8975A
•
•
•
•
•
•
•
•
23