Keysight NFA Series Performance Verification And Calibration Manual page 69

Noise figure analyzers
Hide thumbs Also See for NFA Series:
Table of Contents

Advertisement

Step 1. Open Excel.
Step 2. Open the .csv file from the A: drive. From the
Figure 2-10
Step 3. Increase the column widths to 20 to view the data more accurately.
Step 4. Enter the following formula into cell D1 = SUM(B$1:B1), copy this
Step 5. Enter the following formula into cell D202 =D201/2. This value returns
Performing the frequency accuracy calculations using the CSV
file in a spreadsheet
The following procedure uses Microsoft Excel.
(A:). Select files of
Look in:
Example Excel File
Column A contains the measured frequency. Column B contains the
corresponding Linear P
information.
formula and paste it from cell D2:D201.
the ½ Power reading.
Chapter 2
Text Files ( *.prn; *.txt; *.csv).
Type:
value. Disregard Column C null terminator
hot
Test Descriptions
Frequency Accuracy
menu select
File
Open
and
49