Operator & Service Manual
11.3 REPROGRAMMING AFTER OVERHAUL AND REPAIR PROCEDURES
A flow calibration procedure is required after replacing any of the following components or assemblies: ATF Module, Compres-
sor and Product Tank, PCB Circuit Board, Proportional Valve. If any of these components are replaced, the SAROS must be re-
turned to CAIRE or a factory-authorized service center to have the flow calibration performed before returning to use. Equipment
or procedures to perform a flow calibration are not contained in this manual.
NOTE: Further damage to the SAROS or other components is possible if the flow calibration is not performed.
11.4 SAROS MAIN COMPONENTS
Inlet Cap Assembly
Part # 9756-SEQ
PCB Assembly
Part #9710-SEQ
Compressor/Product
Tank Assembly
Part #SP9240-SEQ
ATF
Assembly
Part # SP9729-SEQ
PCB Wire Harness Assembly
Part # 9201-SEQ
Figure 11-1. Inside SAROS
42 | 9735-SEQ H