Membrane Panel Replacement (9731-Seq); Figure 11-73. Peel; Figure 11-74. Clean; Figure 11-75. Remove Psa Cover - Saros 9400R-SEQ Operators & Service Manual

Table of Contents

Advertisement

Operator & Service Manual

11.8.9 MEMBRANE PANEL REPLACEMENT (9731-SEQ)

11.8.9.1 After performing steps 11.7.1.1 thru 11.7.1.8 from "INLET CAP REMOVAL".
11.8.9.2 Peel off the User Panel (9731-SEQ) as shown in the
figure below.

Figure 11-73. Peel

11.8.9.3. Clean the surface shown in the figure below with Iso-
clean
propyl Alcohol before re-applying the new User Panel.
surface

Figure 11-74. Clean

11.8.9.4 Remove the PSA protective cover from the new User
Panel (9731-SEQ). Insert the Flex Cable as shown in the figure
below.

Figure 11-75. Remove PSA cover

11.8.9.5 Position and apply the new User Panel as shown in the
figures below. Press firmly on the entire surface of the new User
Panel to insure that it is bonded correctly.

Figure 11-76. Apply new User Panel

11.8.9.6 Reconnect the small tube (4056-SEQ [2.3"]).
11.8.9.7 Perform steps 11.8.10.4 - 11.8.10.11 from section "INLET CAP INSTALLATION".
70 | 9735-SEQ H