Compressor Fan Replacement; Proportional Valve Replacement (5293-Seq); Figure 11-57. Fan; Figure 11-58. Compressor Fan Wire - Saros 9400R-SEQ Operators & Service Manual

Table of Contents

Advertisement

Operator & Service Manual
11.8.5.1 Attach the Fan subassembly (9767-SEQ) with the flow
arrows towards the Bulkhead as shown on figure. Apply Loctite
222MS (1405-SEQ) on the screws and torque to 2.0 in-lbs.
11.8.5.2 Connect the Compressor Fan wire to the PCB Assembly
(If PCBA Assembly is connected)
NOTE: A flow calibration procedure is required after replacing <the ATF or Compressor, the PCB Circuit Board, the Proportional Valve>
If this components is replaced, the SAROS must be returned to CAIRE or a factory-authorized service center to have the flow calibra-
tion performed before returning to use.
11.8.6.1 Attach the two screws from the Proportional valve in to the Proportional Block (White) below it and Torque to 3.8 in-
lbs.

11.8.5 COMPRESSOR FAN REPLACEMENT

11.8.6 PROPORTIONAL VALVE REPLACEMENT (5293-SEQ)

proportional
valve block
9231-SEQ

Figure 11-59. Proportional Valve

9913-10-SEQ

Figure 11-57. Fan

fan connector

Figure 11-58. Compressor Fan Wire

screws
9203-7-SEQ
9735-SEQ H | 65
note wire
location
screws
ATF wires go
through slot

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for Saros 9400R-SEQ

Table of Contents