Operator & Service Manual
11.7.2.8 Disconnect the Breath Detect Tube.
disconnect
Figure 11-14. Tube Disconnect
11.7.2.9 Remove the Inlet Cap (Part# 9756-SEQ) and guide AC
Power Cable under the frame of the PCB Assembly.
Guide
Wires
Figure 11-15. Inlet Cap Removal
11.7.2.10 Remove the 3 screws (9912-1) to detach the Top Strut (9778-SEQ).
Figure 11-16. Detach Top Strut
52 | 9735-SEQ H