Table 5-1. Typical New Battery Operating Time - Saros 9400R-SEQ Operators & Service Manual

Table of Contents

Advertisement

Operator & Service Manual
WARNING: IF THE VEHICLE'S DC POWER SOURCE DROPS BELOW 20.0 VOLTS (BROWNOUT CONDITION) THE SAROS
WILL REVERT TO BATTERY OPERATION IF THE BATTERY IS PRESENT.
TYPICAL NEW BATTERY OPERATING TIME
Battery Operating Time: The remaining operating time of the Battery is displayed on the user control panel when in Battery mode.
A variety of factors, such as flow setting, Pulse or Continuous Flow Mode, temperature or age will impact the operating time. The
following table provides operating time estimates for the SAROS operating on a fully charged Battery given certain flow settings
and nominal operating conditions.
CONTINUOUS FLOW
SETTING (LPM)
1.0
2.0
3.0
NOTE: Battery run time will degrade under high temperature use and as the battery cells age.
CAUTION: Store the battery in a cool and dry location. This will help to assure the longevity of the battery.
CAUTION: U.S. Department of Transportation (DOT) and United Nations (U.N.) regulations require the removal of
the battery from the device for all international airline travel when the SAROS is checked as luggage. When ship-
ping the SAROS, the battery must also be removed from the device and packaged properly.
CAUTION: Only replace the CAIRE Inc. battery supplied with the device with a CAIRE Inc. battery.
WARNING: DO NOT ATTEMPT TO OPEN THE BATTERY; THERE ARE NO SERVICEABLE PARTS INSIDE THE BATTERY.
WARNING: DO NOT STORE THE SAROS WITH THE BATTERY INSTALLED IN THE UNIT. RETURN THE BATTERY TO AN
AUTHORIZED SERVICE CENTER OR CAIRE INC. FOR PROPER DISPOSAL.
WARNING: DO NOT DISASSEMBLE, INCINERATE, OR HEAT THE BATTERY ABOVE 140° F (60° C). THE BATTERY, USED
IN THIS DEVICE, MAY PRESENT A RISK OF FIRE OR CHEMICAL BURN IF MISTREATED.
TYPICAL BATTERY RECHARGE TIME
The typical time to recharge your Battery, in order to achieve 80% capacity, from a fully discharged Battery is 1.5 hours, while
operating at 3 LPM continuous flow.
If the Battery becomes too warm during discharging, recharging will not begin until the Battery sufficiently cools. Removing the
Battery and allowing it to cool may expedite this cooling process.
INITIAL BATTERY CHARGING
The new Battery supplied with the SAROS is not fully charged when it is shipped from the factory. Pushing the Test Button on the
Battery will indicate the level of charge within. Before using the SAROS for the first time, fully charge the Battery.
Attach the Battery by aligning the connector on the battery to the mating part on the bottom of the SAROS. The Battery will click
into place. Depress latching pin button prior to connecting battery.
CONTINUOUS FLOW
BATTERY OPERATING TIME
1.1 hours
45 min
30 min

Table 5-1. Typical New Battery Operating Time

PULSE FLOW BATTERY
OPERATING TIME
1.2 hours
1.1 hours
53 min
47 min
46 min
37 min
PULSE FLOW MODE SETTING
(ML)
16
32
48
64
80
96
9735-SEQ H | 27

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents