Saros 9400R-SEQ Operators & Service Manual page 44

Table of Contents

Advertisement

Description
Indicator
Power
Loss of Power
ON
Missing Power
ON
2 beeps of 0.5s
Low Battery
every 119s
Cannot Charge Bat-
2 beeps of 0.5s
tery
every 119s
External voltage too
ON
high
3 beeps after
Low 9 volt battery
POST
Physical / Functional
2 beeps of 0.5s
Key Stuck On
every 119s
Compressor Stall
ON
Compressor Driver
ON
Fault
2 beeps of 0.5s
Pulse Mode Disabled
every 119s
Motor Controller
ON
Fails Post
44 | 9735-SEQ H
Audio
Ext. Power
LED
NC
NC
NC
NC
(1s/1s)
NC
NC
NC
NC
NC
NC
Table 11-2. Alarm Indications and Codes (continued)
Battery
Alarm
Icon
Code
This alarm shall activate during production when the
system determines that valid power has been lost.
NC
4000
O2 Production and battery charging shall stop, and
the electronics shall shut down after 10 seconds.
This alarm shall activate if the user attempts to start
the unit while there is no usable power source con-
NC
4010
nected. O2 Production and battery charging shall not
begin and the electronics shall shut down after 10
seconds.
The alarm shall activate when the time remaining on
(1s/1s)
9540
battery is ≤ 6 minutes. The alarm shall cease when
the time remaining is ≥ 12 minutes.
This alarm shall activate if system is attempting to
(1s/1s)
952C
charge the battery, but can't do so.
This alarm shall activate when the external voltage
is greater than 30V. O2 Production and battery charg-
NC
AC00
ing shall not begin and the electronics shall shut
down after 10 seconds.
NC
None
The 9 volt battery measured less than 6.5V.
This alarm shall activate when a key is active for
NC
8110
greater than 7 seconds.
NC
9194
O2 production shall stop or cannot be started.
NC
9195
O2 production shall stop or cannot be started.
This alarm shall activate when a breath sensor
NC
A200
malfunction is detected. The system shall not enter
pulse mode.
NC
FD00
O2 production shall not be started.
Operator & Service Manual
Description

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for Saros 9400R-SEQ

Table of Contents