Saros Oxygen System Specifications; Table 1-1. Saros Specifications - Saros 9400R-SEQ Operators & Service Manual

Table of Contents

Advertisement

Dimensions
• With battery
• Without battery
Weight
• With Battery
• Without Battery
Flow Settings
• Continuous Flow
• Pulse flow
Continuous Flow Accuracy
Oxygen concentration
Maximum System Pressure
Oxygen Output Pressure
Oxygen Concentration Status Indicator
Nominal Sound Level
Operating Environment
• Temperature
• Humidity
Storage Environment
• Temperature
• Humidity
Nominal Power
3 LPM Continuous Flow
Nominal Battery operating time
at 3 LPM Continuous Flow
Battery Life
Continuous Flow Indication
Audible Alarm Indicators
Back-Up Alarm Power
Filters
Device Classification
12 | 9735-SEQ H

1.2 SAROS OXYGEN SYSTEM SPECIFICATIONS

26.80" long x 4.375" diameter (68.1 cm long x 11.1 cm diameter)
23.25" long x 4.375" diameter (59.1 cm long x 11.1 cm diameter)
12.25 pounds (5.56 kg)
10.00 pounds (4.54 kg)
1, 2, 3 LPM
16, 32, 48, 64, 80, 96 ml
1LPM: +/- 0.4 LPM
2LPM: +/- 10%
3LPM: +/- 10%
93% +/- 3% for all flow settings
15 psig (204.7 kPa)
4.0 psig (128.9 kPa) nominal
Green Light = Normal Operation
Yellow Light Flashing = Warning or Caution, less than 85% ± 3%
Red Light = Abnormal Operation, less than 70% ± 5%
<59 dB(A)
32º F to 109º F (0° to 43°C)
10% to 95%, Non-condensing, 82.4°F (28°C) Maximum Dew point
-4º F to 140º F (-20° to 60°C)
Up to 95% Non-condensing
≤130 Watts
30 Minutes
80% Nominal Capacity after 200 Charge/Discharge cycles
Expressed in liters per minute (LPM)
• Loss of Power/Hot Battery
• Low Battery/Warm Battery
• Low Oxygen Output
• No breath detection
• O2 Flow Outside Normal Limits
• Low 9V battery
• Unit Malfunction
9V Internal Battery
Air Inlet Filter, HEPA and Exhaust Filter
IEC Class I, Type BF Applied Part, IPX1

Table 1-1. SAROS Specifications

Operator & Service Manual

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents