Operator & Service Manual
11.7.2.4 Cut green cable tie and disconnect the Oxygen Outlet
Tube from the Flow Valve.
oxygen
outlet
tube

Figure 11-10. Oxygen Outlet Tube Disconnect.

11.7.2.5 Remove tape (Part# 2696-SEQ) from Front Panel Flex
Cable and Top Mounting Struts (Part# 9778-SEQ). Lift up lock-
ing connector and disconnect the Front Panel Flex cable.
front panel
flex cable

Figure 11-11. Front Panel Flex Cable Disconnect

11.7.2.6 Disconnect the Ambient Air Thermistor Connector and
9V Battery Connector.
9V Battery
connector
ambient
thermistor
connector

Figure 11-12. Thermistor Connector and 9V Battery Connector

11.7.2.7 Loosen the screws for the AC Power Connector and
remove the wires.
loosen

Figure 11-13. Inlet Cap Power

9735-SEQ H | 51