Appendix A A-1. Saros Performance Verification Checklist - Saros 9400R-SEQ Operators & Service Manual

Table of Contents

Advertisement

Operator & Service Manual
SAROS Serial Number:
Hours of Operation:
A. General Operation Test
1) Battery installs properly
2) External power cord (AC or DC)
connects securely
3) Unit powers on after holding ON/OFF
button for 3 seconds
4) Audible alarm sounds upon start-up
5) 5-minute warm up period complete with
no audible alarm
6) Utility Mode shows no active alarms
after 5-minute warm up
7) Battery icon visible on screen
8) Battery is charging or already fully
charged
9) Blocked Flow Alarm is functional
10) Increase and decrease flow buttons are
functional
11) Pulse flow mode button is functional
12) No breath alarm is functional
13) Unit runs on battery power
14) No battery alarm is functional
15) Automatic re-start when power is
reconnected
B. Flow and O
Concentration Test
2
1) Continuous Flow Setting Verification
C3 (3.0 LPM) ) O
Concentration
2
C3 (3.0 LPM) flow -- 60 second average
reading
1
C2 (2.0 LPM) ) O
Concentration
2
C2 (2.0 LPM) flow -- 60 second average
reading
1
2) Pulse Flow Mode Verification
P48 (48 mL) -- 5 bolus average reading
P96 (96 mL) -- 5 bolus average reading
C. Battery Run Time Test
C3 (3.0 LPM)
Footnotes: (1) The SAROS is BTP compensated. If measuring concentration and/or flow with a device that is non-Argon compensated and
non-BTP compensated, the acceptable range of the device readings will be reduced by 0.1 LPM.
Technician Name (print): _______________________________________
Technician Signature: _________________________________________
APPENDIX A
A.1 SAROS PERFORMANCE VERIFICATION CHECKLIST
Pass
Acceptable Range
90 - 96 %
Argon-Compensat-
Non Argon-Compen-
ed Testing Device:
sated Testing Device:
2.70 - 3.30 LPM
2.60 - 3.20 LPM
90 - 96 %
Argon-Compensat-
Non Argon-Compen-
ed Testing Device:
sated Testing Device:
1.80 - 2.20 LPM
1.70 - 2.10 LPM
40.8 - 55.2 mL
81.6 - 110.4 mL
Acceptable Range
> 24 min
Table A-1. SAROS Performance Verification Checklist
Date of Testing:
1:
2:
1:
2:
Date: _______________________
Fail
Recorded Reading
3:
4:
5:
Ave:
3:
4:
5:
Ave:
Recorded Reading
9735-SEQ H | 77