Operator & Service Manual
11.8.2.4 Apply cap (1137-SEQ) to Product Tank outlet fitting to
prevent ATF desaturation!
11.8.2.5 Apply 2 Cable Ties as shown on figure.
NOTE: Pre Loop Zip Ties.
11.8.2.6 Attach the Exhaust Filter, Tube, Muffler Assembly and Tube, Muffler as shown in figure below.
NOTE: Ensure Arrow on Filter points in the correct direction. Install Cable Ties (6968-SEQ) as shown in figure below. Ensure the com-
pressor exhaust elbow tubing is through the bulkhead before/during assembly.
While applying Kapton tape, pull muffler tube to establish clearance. Make sure that tube, muffler, and muffler tube are laying
down on ATF surface.
NOTE: Pulling too hard on the Exhaust Tube may cause the tube to tear.
exhaust filter
(9221-SEQ)
v
clearance
tube
2x3" kapton tape
(9227-SEQ)
(2696-SEQ)
Figure 11-47. Open SAROS
cap off
fitting
Figure 11-45. Cap
Figure 11-46. Cable Ties
muffler assy
(9314-SEQ)
muffler tube
(9390-SEQ)
v
pull
9735-SEQ H | 61