Saros 9400R-SEQ Operators & Service Manual page 46

Table of Contents

Advertisement

Audio
Description
Indicator
Temp
ATF Temperature
ON
Sensor Failure
Compressor Too
ON
Hot
Compressor Tem-
perature Sensor
ON
Failure
Compressor and
ATF Temperature
ON
Sensor Failure
Power Board Too
ON
Hot
2 beeps of
Power Board
0.5s every
Thermistor Failure
119s
Motor Driver
ON
Board Too Hot
Motor Driver
2 beeps of
Board Thermistor
0.5s every
Failure
119s
Battery Too Hot
ON
2 beeps of
Battery Thermistor
0.5s every
failure
119s
2 beeps of
Battery Chip
0.5s every
Thermistor failure
119s
Battery Too Cold
OFF
to Use
Battery Too Cold to
OFF
Charge
Product Tempera-
ON
ture Sensor Failure
Battery Too Hot to
OFF
Charge
Product and Ambi-
ent Temperature
ON
Sensor Failure
46 | 9735-SEQ H
Ext. Power
Battery
LED
Icon
NC
NC
8900
NC
NC
9100
NC
NC
9110
NC
NC
9110
NC
NC
9200
NC
NC
9210
NC
NC
9300
NC
NC
9310
NC
(1s/1s)
9510
NC
(1s/1s)
9550
NC
(1s/1s)
9560
No Water-
NC
95C0
fall
No Water-
NC
95CC
fall
NC
NC
A410
A410
No Water-
NC
None
fall
A410
NC
NC
AB10
Table 11-2. Alarm Indications and Codes (continued)
Alarm
Code
O2 production shall continue.
O2 production shall stop or cannot be started.
O2 production shall continue.
O2 production shall stop or not be started depending on
Compressor Temperature Sensor and ATF Temperature
Sensor condition.
O2 production and battery charging shall stop or cannot
be started. The electronics shall shut down after 10
seconds when on battery power. The display shall be
blanked after 10 seconds when on external power.
Production shall continue. Battery charging if required
shall continue.
O2 production and battery charging shall stop or cannot
be started.
Production shall continue. Battery charging if required
shall continue.
Battery shall not be charged or discharged.
The system shall not use the battery if the independent
battery thermistor determined to be open or shorted.
The system shall not use the battery if the reported bat-
tery chip thermistor is determined to be open or shorted.
The battery shall not be charged or discharged.
The battery shall not be charged.
O2 production shall continue.
Product Temperature Sensor and Ambient Temperature
Sensor condition.
The battery shall not be charged.
O2 production shall stop or not be started depending on
Product Temperature Sensor and Ambient Temperature
Sensor condition.
Operator & Service Manual
Description

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for Saros 9400R-SEQ