11.8.10.6 Insert Front Panel Flex cable flush with locking con-
nector. Apply 2" of Kapton Tape (2696-SEQ) on the Mounting
STRUT.
11.8.10.7 Connect the Ambient Air Thermistor and connect 9V
Battery Connector as shown.
11.8.10.8 Attach the Oxygen Outlet Tube (6981-SEQ) to the
Flow Valve outlet port and secure with a cable tie.
11.8.10.9 Secure the three Inlet Cap Mounting screws (9914-6-
SEQ) to each STRUT.
72 | 9735-SEQ H
front panel
flex cable
securing tape
(2696-SEQ)

Figure 11-81. Flex cable

9v battery
connector
Ambient Air
Thermistor

Figure 11-82. Ambient Air Thermistor

oxygen
tube

Figure 11-83. Oxygen Outlet Tube

inlet cap
mounting screws
9914-6-SEQ

Figure 11-84. Inlet Cap mounting screws

Operator & Service Manual