Disk Drive Operations

Load and Store

Load and Store
When you choose Load or Store, you next need to set the field immediately
to the right. This field presents at least four choices: All, System, Analyzer,
and Scope. If you have other software loaded, it might add to the list of
choices.
All
Choose All to store or load system, analyzer, and scope information. If you
are storing, three files (one for the system, one for the analyzer, and one for
the scope) are created. The system file ends in "._ _", the analyzer file in
"._A", and the scope file in "._B".
System
System files store system configurations. System information for the
HP 1660CS-series consists of settings for printer, controller, RS-232-C,
HP-IB, shade, and sound. LAN settings are not saved. System configuration
files end in two underscores and have a file type of 16[6/7]x_config.
Analyzer
Analyzer configuration files store measurement setups, including data. If you
are storing the current measurement and an inverse assembler is already
loaded, when you reload the file you are now creating it will try to pull in the
inverse assembler. Other attributes stored in analyzer configuration files
include labels, trigger sequence, arming configuration, measurement data,
markers, analyzer names, and pod assignments. Analyzer configuration files
end in "._A" and have a file type of 166xan_config.
Scope
Oscilloscope configuration files store measurement setups, including data.
Attributes stored in scope configuration files include labels, trigger sequence,
arming configuration, measurement data, markers, and channel assignments.
Oscilloscope configuration files end in "._B" and have a file type of
166Xsc_config.