Analyzer Problems

This section lists general problems that you might encounter while
using the analyzer.
Intermittent data errors
This problem is usually caused by poor connections, incorrect signal levels, or
marginal timing.
With the logic analyzer and all connected equipment turned off, remove
and reseat all cables and probes; ensure that there are no bent pins on
the preprocessor interface or poor probe connections.
Adjust the threshold level of the data pod in the Format menu to match
the logic levels in the system under test.
Use an oscilloscope to check the signal integrity of the data lines.
Clock signals for the state analyzer must meet particular pulse shape and
timing requirements. Data inputs for the analyzer must meet pulse shape and
setup and hold time requirements.
See Also
See "Capacitive Loading" in this section for information on other sources of
intermittent data errors.

Unwanted triggers

Unwanted triggers can be caused by instructions that were fetched but not
executed.
Add the prefetch queue or pipeline depth to the trigger address.
The depth of the prefetch queue depends on the processor that you are
analyzing. Suppose you are analyzing a pipelined processor having fetch,
decode, execute, and memory stages. The processor fetches 32-bit words. To
ensure that the processor has begun executing a particular routine when the