Cross-Arming Trigger Examples - HP 1660CS-Series User Manual

Logic analyzers
Table of Contents

Advertisement

Cross-Arming Trigger Examples

The following examples use cross arming to coordinate measurements
between two separate analyzers within the logic analyzer or between
analyzers and the oscilloscope. The analyzers can be configured as
either a state analyzer and timing analyzer, or two state analyzers. It
is not possible to set both to timing.
You set up cross arming in the Arming Control menu (obtained by
selecting Arming Control in the Trigger menu). When coordinating
measurements between two instruments, you need to select Count
Time to correlate the measurements. When correlating
measurements between an analyzer and the scope you also need to
set the oscilloscope Trigger Mode to Immediate.
"Arming and Additional Instruments" in Chapter 4 for more
See also
information on cross arming.