Model 8902A
Operation
Store/Recall
FUNCTIONS
STORE. RECALL
DESCRIPTION
The Measuring Receiver can store a complete instrument stat.e (including all front-panel functions
and most Special Functions) in any of its eight storage registers. The stored data is retained in non-
volatile memory when the instrument is
o f f
even when the power cord is removed. Any of these
stored states can then be re-initiated.
PROCEDURE
To store the current instrument state, select STORE (Blue Key,
7). Then key in a register number
( 1
through 8).
To recall a previously-stored instrument state, select RECALL (Blue Key, 8). Then key
in the desired
register number (1 through 8).
HP-IB PROGRAM CODES
RECALL
=
RC
STORE
=
TR
COMMENTS
Regarding storage of
RF
power calibration factors:
.The current reference calibration factor and the frequency/calibration-factor pair can be stored.
.Manual calibration factors can be stored.
.The table of calibration factors cannot be stored with this function, but two tables of calibration
factors can be separately stored in the instrument's non-volatile memory. (Refer to RF Power under
"Comments" on page 3-83.)
3-253