To interleave trace lists

Set up both analyzers as state analyzers.
1
Go to the Trigger menu of the first analyzer.
2
Set Count to Time, and set up the trigger.
3
The logic analyzer uses the time tags stored with each state to determine the
ordering of states shown in an interleaved trace list.
Set Count to Time, and set up the trigger on the second analyzer.
4
The second analyzer does not need to be run from the first analyzer, but both
analyzers must have "Count Time" turned on in order to correlate the data.
Make the measurement run.
5
Go to one of the Listing menus.
6
Select one of the label fields in the trace list display, then select
7
Interleave.
Select the name of the other analyzer and the label to interleave.
8
Interleaved data is displayed in a light shade. Trace list line numbers of
interleaved data are indented. The labels identifying the interleaved data are
shown above the labels for the current analyzer, and are displayed in a light
shade.
If you have problems with the procedure, check that each analyzer has an
independent clock from the target system.
Interleaved trace lists on the HP 1661CS
Special displays
To interleave trace lists
5-31