Preparing For Use

To degauss the display

To degauss the display
If the logic analyzer has been subjected to strong magnetic fields, the
CRT might become magnetized and display data might become
distorted. To correct this condition, degauss the CRT with a
conventional external television type degaussing coil.

To clean the logic analyzer

With the instrument turned off and unplugged, use mild soap and
water to clean the front and cabinet of the logic analyzer. Harsh soap
might damage the water-base paint. Do not immerse the logic
analyzer in water.

To test the logic analyzer

If you require a test to verify the specifications, the HP 1660C/CS
Series Logic Analyzers Service Guide is required. Start at the
beginning of chapter 3, "Testing Performance."
If you require a test to initially accept the operation, perform the
self-tests described in Troubleshooting in this chapter.
If the logic analyzer does not operate correctly, go to the flow charts
provided in Troubleshooting in this chapter.