Calibrating and Adjusting
To calibrate the oscilloscope
Load the Default Calibration Factors
Note that once the default calibration factors are loaded, all calibrations must be
done. This includes all of the calibrations in the Self Cal menu. The calibration must
be performed in the exact sequence listed below.
1 Press the System key. Select System, then select Scope.
2 Select the menu field (top row, second from the left), then select Scope Calibration
from the pop-up menu.
3 Select the Mode field, then select Service Cal from the pop-up menu.
4 Select the Procedure field, then select Default Values from the pop-up menu.
5 Select the Start field and follow the instructions on the display.
~(
Scope Calibration
) (AutoscOle)
~ ~
Mode
ServIce Col
Stort
FUUCTIOll
'"
GOIn
Offset
H~stereSl~
Tngger
De10y
Tir~e
lIull
Log 1C Tn gger
I"roc~dure
Default'lalu\!s
Continue
I
C I
C2
p •
p~ss\!d,
f
•
I oi I eo. D • deloulled. C • C\lrrupted
Prell>;ed by.
If
new solllMre reVISion mtnout
n-col
After you select the Start field, you can abort the calibration procedure by selecting either the
Mode or Procedure fields
if
the Continue field is still displayed on the screen.
4-4