Keysight N437 Series Programmer's Manual page 68

Lightwave component analyzer
Table of Contents

Advertisement

1
Remote Operation
68
C#
(property) LaserWavelength_nm
Example
:SOUR:WAV 1550
Command
:SOURce:WAVelength:ALL?
Syntax
:SOURce:WAVelength:ALL?
Description
Retrieves a list showing all available wavelengths of the LCA test head.
Parameters
none
Response
<string> comma separated list of wavelengths units.
C#
(property) SourceWvl
Example
:SOUR:WAV:ALL? -> Wvl_1310nm, Wvl_1550nm
Command
:THEAd:INPut:MODe?
Syntax
:THEAd:INPut:MODe?
Description
Gets the current optical input of the LCA test-head.
Parameters
none
Response
<string> Standard | HighPower
C#
(property) OpticalInput
Example
THEA:INP:MODE? -> Standard
Command
:THEAd:INPut:MODe
Syntax
:THEAd:INPut:MODe<wsp>STAN|HIGH description:Sets the current
optical input of the LCA test-head. parameters:<string> HIGH | STANdard
Response
none
C#
(property) OpticalInput
Example
THEA:INP:MODE HIGH
Keysight N437x Series Lightwave Component Analyzer, Programmer's Guide