Keysight N437x Series Lightwave Component Analyzer, Programmer's Guide
Command
:SOURce:STATe?
Syntax
:SOURce:STATe?
Description
Gets the current state of the LCA optical output.
Parameters
none
Response
<string> LaserOn | LaserOff
C#
(property) LaserState
Example
:SOUR:STAT? -> LaserOn
Command
:SOURce:STATe
Syntax
:SOURce:STATe<wsp>ON|OFF|1|0 description:Sets the current state of the
LCA optical output. parameters:<string> ON | 1 | OFF | 0
Response
none
C#
(property) LaserState
Example
:SOUR:STAT ON
Command
:SOURce:WAVelength?
Syntax
:SOURce:WAVelength?
Description
Gets the current wavelength of the LCA optical output.
Parameters
none
Response
<string> The wavelength as string together with the unit.
C#
(property) LaserWavelength_nm
Example
:SOUR:WAV? -> <Wvl_1550nm
Command
:SOURce:WAVelength
Syntax
:SOURce:WAVelength<wsp><wavelength>
Description
Sets the current wavelength of the LCA optical output. The available
wavelengths can be retrieved with :SOUR:WAV:ALL?
Parameters
wavelength <string> the wavelength value as string, unit is nm e. g. 1550.
Response
none
Remote Operation
1