Keysight N437 Series Programmer's Manual page 26

Lightwave component analyzer
Table of Contents

Advertisement

1
Remote Operation
Property
UseOpticalConnData
SrcAttOpt_dB
RcvAttOpt_dB
SrcRefIdx
RcvRefIdx
SrcLengthOpt_m
RcvLengthOpt_m
UseOpticalS2PFile
OptRcvFile
OptSrcFile
26
The following properties control additional optical path deembedding.
Description
With this property you could switch the whole optical path
deembedding on or off.
Specify the optical attenuation on the source path in dB. In default
mode only evaluated for O/E and O/O measurements.
Specify the optical attenuation on the receiver path in dB In default
mode only evaluated for E/O and O/O measurements.
Specify the refractive index of the source path in dB.
In default mode only evaluated for O/E and O/O measurements.
Specify the refractive index of the receiver path in dB.
In default mode only evaluated for E/O and O/O measurements.
Specify the geometrical length of the source path in m.
In default mode only evaluated for O/E and O/O measurements.
Specify the geometrical length of the receiver path in m.
In default mode only evaluated for E/O and O/O measurements.
Specify if you want to describe the optical paths by the parameters
above or by transmission data stored in a s2p file. Only the S21
transmission data is used.
The name of the s2p file to use for additional adaptor deembedding
on the receiver side
In default mode only evaluated for E/O and O/O measurements.
The name of the s2p file to use for additional adaptor deembedding
on the source side.
In default mode only evaluated for O/E and O/O measurements.
Keysight N437x Series Lightwave Component Analyzer, Programmer's Guide
Type
Default value
Boolean
False
Double
0.0
Double
0.0
Double
0.0
Double
0.0
Double
0.0
Double
0.0
Boolean
False
String
Empty string
String
Empty string

Advertisement

Table of Contents
loading

Table of Contents