Keysight N437 Series Programmer's Manual page 62

Lightwave component analyzer
Table of Contents

Advertisement

1
Remote Operation
62
Command
:PARameter:SOURce:REFR:INDex
Syntax
:PARameter:SOURce:REFR:INDex<wsp><index>
Description
Specifies the refractive index of the source path in dB. In default mode only
evaluated for O/E and O/O measurements.
Parameters
index <double> the refractive index value in dB
Response
none
C#
(property) SrcRefIdx
Example
:PAR:SOUR:REFR:IND 0.13
Command
:PARameter:USER:CALIbration:CALSet?
Syntax
:PARameter:USER:CALIbration:CALSet?
Description
Retrieves the name of a Calset on the network analyzer to be used for the
user calibration measurement. If an empty string is returned, the current
Calset is used. If "NONE" is returned, no Calset is applied for the user
calibration measurement.
Parameters
none
Response
<string> NONE | path and file name
C#
(property) CalSetUserCal
Example
:PAR:USER:CAL:CALS? -> c:\temp\calset1.s2p
Command
:PARameter:USER:CALIbration:CALSet
Syntax
:PARameter:USER:CALIbration:CALSet<wsp>[NONE | "<path string>"]
Description
Specifies the name of a Calset on the network analyzer to be used for the
user calibration measurement. If an empty string is passed, the current
Calset is used. If "NONE" is passed, no Calset is applied for the user
calibration measurement.
Parameters
NONE| <string> |No argument, None or the path and file name surrounded
by double quotes.
Response
none
C#
(property) CalSetUserCal
Example
:PAR:USER:CAL:CALS "c:\temp\calset1.s2p"