Keysight N437 Series Programmer's Manual page 54

Lightwave component analyzer
Table of Contents

Advertisement

1
Remote Operation
54
Response
<double> power value, the default unit is dBm.
C#
(property) OpticalPower_dBm
Example
:PAR:OPT:OUT:POWer -> -1
Command
:PARameter:OPTical:OUTput:POWer
Syntax
:PARameter:OPTical:OUTput:POWer<ws><power> description:Specify the
optical output power of the LCA in dBm.
Parameters
power <double>, power value in dBm
Response
none
C#
(property) OpticalPower_dBm
Example
:PAR:OPT:OUT:POWer -1
Command
:PARameter:OPTical:PATH:DEEMbedding?
Syntax
:PARameter:OPTical:PATH:DEEMbedding?
Description
Returns whether the whole optical path de-embedding is set on or off.
Parameters
none
Response
<string> 1 = optical path de-embedding is enabled, 0 = disabled
C#
(property) UseOpticalConnData
Example
:PAR:OPT:PATH:DEEM? -> 1
Command
:PARameter:OPTical:PATH:DEEMbedding
Syntax
:PARameter:OPTical:PATH:DEEMbedding<wsp>ON|1|OFF|0
Description
Switches the whole optical path de-embedding on or off.
Parameters
<string> ON | 1 = enabled optical path de-embedding, OFF | 0 = disable
Response
None
C#
(property) UseOpticalConnData
Example
:PAR:OPT:PATH:DEEM? -> 1
Keysight N437x Series Lightwave Component Analyzer, Programmer's Guide

Advertisement

Table of Contents
loading

Table of Contents