Keysight N437 Series Programmer's Manual page 60

Lightwave component analyzer
Table of Contents

Advertisement

1
Remote Operation
60
Command
:PARameter:SOURce:EXTernal?
Syntax
:PARameter:SOURce:EXTernal?
Description
Retrieves the property which defines whether the optional external input is
used or the internal laser of the optical test set is used.
Parameters
None
Response
<string> 0 | 1
C#
(property) LCAMeasParams.ExternalInput
Example
:PAR:SOUR:EXT? ? 1
Command
:PARameter:SOURce:POWer:STATe?
Syntax
:PARameter:SOURce:POWer:STATe?
Description
Retrieves the internal laser state, on or off. Note: the value of this property
is only evaluated in advanced mode. In default mode the laser is switched
on or off according to the measurement type.
Parameters
none
Response
<string> 1 = internal laser is on, 0 = internal laser is off
C#
(property) Laser_On
Example
:PAR:SOUR:POW:STAT? -> 1
Command
:PARameter:SOURce:POWer:STATe
Syntax
:PARameter:SOURce:POWer:STATe<wsp>ON|1|OFF|0
Description
Switches the internal laser on or off. Note: The value of this property is only
evaluated in advanced mode. In default mode the laser is switched on or
off according to the measurement type
Parameters
<string> ON | 1 to switch the laser on, OFF | 0 to switch the laser off
Response
none
C#
(property) Laser_On
Example
:PAR:SOUR:POW:STAT ON