VS Series Helium Mass Spectrometer Leak Detectors
The commands listed in Table A-2 are used to set non-volatile operating parameters. The
current value of the operating parameter is changed to the new value. If the leak detector is
in back panel control, the response is cant.
Command
INIT-AZ<0
INIT-CL-XFER
INIT-CPLOCKOUT
INIT-DATE
INIT-DISPLAY
INIT-EMISSION
INIT-EXPONENT
INIT-EXTLEAK
INIT-FILAMENT
INIT-GAIN
A-8
Table A-2 Non-Volatile Operating Parameters
Preceded by 0 or 1, sets the status of Auto-zero < 0. 0 = off, 1 = on.
Preceded by X.XE-X, sets the value of the contra-flow mode crossover
pressure in Torr.
Sets the status of the control panel lockout. Preceded by (0) or (1), 0 =
OFF, 1 = ON.
Sets the system date and time. Preceded by: dow, mm dd yyyy hh mm ss
AmPm 24-12 mode, Where 2 is the day of the week (dow) (0 = Sunday, 1
= Monday, etc), mm/dd/yyyy (date), hh:mm:ss (time) AmPm (0=AM,
1=PM)24-12 mode (0 for 24 hr, 1 for 12 hr). Examples: 2 02 12 2005 04 16
33 1 0 INIT-DATE(Tuesday, February 12, 2005, 4:16:33 on a 24 hr clock)2
02 12 2005 04 16 33 1 1 INIT-DATE(Tuesday, February 12, 2005, 4:16:33
PM)?CURRDATE reports the stored value
Sets the status of the bar graph display to either a log or linear display.
Preceded by (1) Log or (0) Linear.
Sets the emission of the ion source. Preceded by a three-digit number
(micro-amps), in the range 60 to 165.
Sets the range that the leak detector will operate in when Manual Range is
ON. Preceded by 0 to -11.
Sets the value of the external leak. Preceded by a two-digit leak rate
mantissa with a decimal point after the first digit followed by E-, and then
by a two-digit leak range exponent: the helium leak rate number in atm
cc/sec of the calibration standard leak.
Example: 3.2E-08 INIT-EXTLEAK
Sets the operating filament in the ion source. Preceded by a single digit,
either 1 or 2.
Sets the value of the gain used for adjusting the helium signal to match a
calibration standard leak. Preceded by a two-digit number with a decimal
point after the first digit, in the range 1.0 to 6.0.
Note: Changing Gain changes Zero by the same ratio.