3.1.8.3 Internal Calibrated Leak

Press Internal Calibrated Leak.
The Internal Calibrated Leak screen appears (Figure 3-17).
To use this screen:
1. Press a button to activate its field.
A keypad appears.
2. Enter the required values as shown on the calibrated leak label.
3. Repeat step 1 and 2 for each desired item.
4. Press Done to exit the screen.
The value of the internal calibrated leak, as shown on the CALIBRATION SET-UP and
CONTROL PANEL screens, is compensated for temperature using the calibrated leak
temperature sensor. The calibrated leak is not compensated for temperature if any of the
following conditions exist:
The temperature sensor on the calibrated leak fails or there is a cable connection
problem.
The temperature sensor reading is out of the operating range. The operating
temperature range is from 10.0 to 45.0° C (50 to 113° F).
The internal calibrated leak is not a E-7 or E-8 range leak.