Agilent Technologies VS Series Operation Manual page 137

Helium mass spectrometer leak detectors
Hide thumbs Also See for VS Series:
Table of Contents

Advertisement

VS Series Helium Mass Spectrometer Leak Detectors
Pin
No.
Signal Name
2
ZERO_OUT
3
HOLD_OUT O
4
VENT_OUT
5
START_OUT
6
READY_
OUT
7
TEST_OUT
8
REJECT1_
OUT
9
REJECT2_
OUT
10
REJECT3_
OUT
Table A-6 I/O Connector Pin Signals (Continued)
I/O
Signal
O
Level
Active HIGH when the leak
detector is in ZEROING
mode.
Level
Active HIGH when the leak
detector is in HOLD mode.
O
Level
Active HIGH when the leak
detector is VENT mode.
O
Level
Active HIGH when the leak
detector is in
START/ROUGH mode.
O
Level
Active HIGH when the leak
detector is in READY mode.
O
Level
Active HIGH when the leak
detector is in FINE TEST
mode.
O
Level
Active HIGH when the leak
rate is over set point #1.
O
Level
Active HIGH when the leak
rate is over set point #2.
O
Level
Active HIGH when the leak
rate is over set point #3.
Description
Leak Detector Behavior
Performs the Zeroing
routine.
Performs a HOLD routine
and isolates the test port and
any test objects from the
leak detector vacuum
system.
Performs a VENT routine,
isolates the test port from the
leak detector vacuum system
and vents the test port to
atmosphere.
Performs a ROUGHING
routine on the test port and
any attached test objects.
Indicates that the leak
detector is completing a
STARTING routine and that
the filament and turbo are
OK.
Indicates that the leak
detector is ready for fine leak
testing.
Indicates that the value of
the leak rate exceeds the
value set in the 1REJECT
variable.
Indicates that the value of
the leak rate exceeds the
value set in the 2REJECT
variable.
Indicates that the value of
the leak rate exceeds the
value set in the 3REJECT
variable.
A-15