3.1.4 Output Control

Press Output Control and the Output Control screen appears (Figure 3-8).
Use this screen to select the desired leak rate output, and Bar Graph display mode.
Press to select the
Analog Output type.
Analog Output
Bar Graph
0
1
2
3
4
OUTPUT CONTROL
ANALOG OUTPUT
LINEAR 10V/DEC
LOG 1V/DEC
Figure 3-8 Output Control
Configures the analog outputs:
Linear 10 V/DEC
Linear output running from 0 to 10 V for each decade.
Figure 3-9 shows the linear output voltage as a function
of leak rate.
Log 1 V/DEC
Log output at 1 V per decade. Figure 3-10 shows the log
output voltage as a function of pressure.
For added accuracy add .06 to V
NOTE
Leak Rate = 10
Configures the bar graph:
Linear Only
The linear bar graph leak rate appears across the top of
all screens when selected. In the Linear Bar Graph display
mode, the bar graph display represents the mantissa of the
leak rate, spanning from 0 to 10. The numerical display
(located to the upper right of the bar graph) indicates the
leak rate exponent value. This mode offers exceptional
resolution within a decade.
5
6
7
8
9
BAR GRAPH
LINEAR
LOG
LOG+LINEAR
Vout+ .06
11
/10
.
10
-
10
Press to select the
bar graph units.
DONE
.
out
3-13