Table 4-3 lists the fuses and circuit breakers.
Function
Circuit Breaker
Fuses (PCB overcurrent
protection)
4-4
Always perform a Full Internal Calibration after any
NOTE
spectrometer maintenance:
Spectrometer header replacement
Filament replacement
Ion source assembly replacement or cleaning
Table 4-3 Fuses and Circuit Breakers
Type, MFG, PIN
Main Power Supply: 20 A, 250 V
AC, SCHURTER
Model EF12, PIN
EF12.2196.1110.01
Internal Vacuum Pump: 6 A,
250 V E-T-A, PIN
1110-Fl12-PIMl-6 A
1.85 A, 33 VDC, Raychem
SMD185-2
5x20mm, T4A, 4A, 250V AC,
Cooper/Bussmann S505-4A
Only a qualified technician should replace circuit breakers and
NOTE
fuses. Use only circuit breakers and fuses of the same type and
rating.
Location
Power Entry Module
Rear Panel
Motherboard (RT400)
Valve Driver (Fl)
Ion Source (Fl)
Front Panel Controller
(RT400)
Turbo Controller (F2)