Agilent Technologies VS Series Operation Manual page 138

Helium mass spectrometer leak detectors
Hide thumbs Also See for VS Series:
Table of Contents

Advertisement

VS Series Helium Mass Spectrometer Leak Detectors
Pin
No.
Signal Name
11
REJECT4_
OUT
12
CAL_OUT_
OK
13
AUTO_
MANUAL_
RANGE_IN
14
CAL_IN
16
ZERO_IN
18
HOLD_IN
20
VENT_IN
22
TEST_IN
A-16
Table A-6 I/O Connector Pin Signals (Continued)
I/O
Signal
O
Level
Active HIGH when the test
port pressure is over set
point #4.
O
Level
Active HIGH when
calibration is successful.
I
Level
Active HIGH when Manual
Ranging is selected.
I
> 200
Active HIGH pulse.
ms Pulse
I
> 200
Active HIGH pulse.
ms Pulse
I
> 200
Active HIGH pulse.
ms Pulse
I
> 200
Active HIGH pulse.
ms Pulse
I
> 200
Active HIGH pulse.
ms Pulse
Description
Leak Detector Behavior
Indicates that the value of
the test port pressure
exceeds the value set in the
4REJECT variable.
Indicates the calibration has
successfully completed. Set
low on power-up or if
calibration is not
successfully completed.
Activation causes the leak
detector to change from
AUTO to MANUAL ranging
; Default is AUTO.
Initiate CALIBRATING
routine with Internal
Calibrated Leak; Zeroing
routine is included.
Initiate ZEROING routine.
Initiate HOLD routine and
isolate the test port and any
test objects from leak
detector vacuum system.
Initiate VENT routine and
vent the test port to
atmosphere.
Initiate ROUGHING routine
on the test port and any
attached test objects.