VS Series Helium Mass Spectrometer Leak Detectors
3.1.11 Wireless
When the wireless option is installed into the leak detector, a wireless icon button appears
on the second Setup screen (Menu/Setup/
Press the Wireless icon
❑
appears (Figure 3-23).
Use this screen to assign the RF channel that the remote handheld device uses for:
Communications with the leak detector,
❑
Control of the leak detector
❑
And to get other information relating to wireless communications.
❑
Opens a keypad
to configure the
channel for use.
MAC Address
SYNC Channel
SS ID
0
1
2
3
4
MAC ADDRESS: 00:50:67:11:30:OC
SYNC CHANNEL: 007
SS ID: 086
07
CHANNEL
00:50:67:10:2A:OD
DONE
Figure 3-23 Wireless
Two MAC addresses appear on this screen:
The MAC address at the top of the screen is the leak detector
❑
MAC address.
The MAC address above the hand held button is the MAC
❑
address of the hand held device.
For informational and factory debug purposes only.
Displays the SS ID number, factory set to 086, for informational and
factory debug purposes only.
).
on the Setup screen and the wireless screen
5
6
7
8
9
Ends the session with the current wireless remote.
Any wireless is now free to establish communications
with the VS unit (
Button only appears if a handheld
).
unit is connected
C
10
Turns all access to
the VS unit ON or
✔
OFF.
✕
DONE
DONE
3-27