VS Series Helium Mass Spectrometer Leak Detectors

2.2.4.6 System Information

Press System Information on the Menus screen.
The System Information screen appears (Figure 2-15).
Details of the current leak detector setup and operating conditions are displayed.
The left column displays filament status, reject set point status, turbo status, auto
sequencer status, and audio set point status. The right column displays the
calibrated leak setting, leak rate range settings, and the sensitivity configuration.
Table 2-6 describes the displayed system information in detail.
A blue check
denotes a good
filament; red is bad
.
Displays the Gain
a
Condition
Audio Set Point
See Section 2.2.4.2 "Reject and
Audio Set Points" on page 2-19.
Auto Sequencer
See Section 2.2.4.3 "Auto
Sequencer" on page 2-22.
Calibration Set-Up
See Section 2.2.4.1 "Calibration
Set-Up" on page 2-17.
2-30
0
1
2
3
4
SYSTEM INFORMATION
FILAMENT 1
GAIN: 1.5
CPU: LD01.00.02
LCD: DDO1.00.02
WWW.AGILENT.COM/
Figure 2-15 System Information
Table 2-6 System Information Screen Conditions
Audio Set Point:
Active
Audio Set Point:
Inactive
Auto Sequencer: On
Auto Sequencer: Off
Int Cal Leak
Ext Cal Leak
5
6
7
8
9
INTERNAL CAL LEAK
AUTO RANGING
Display
Audio Set Point is active.
Audio Set Point is inactive.
Auto Sequencer (automatic test cycle)
is enabled.
Auto Sequencer (automatic test cycle)
is disabled.
System is set to calibrate to the internal
standard leak.
System is set up to calibrate to an
external standard leak.
10
-
10
DONE
Description