On-Off Logic - Honeywell Vibrex 2000 User Manual

Digital dynamic balancer / analyzer systems
Table of Contents

Advertisement

Chapter 6 – Maintenance
6.2.4.2
Once the on-off logic senses a press of the
powers up the battery-driven 12V power supply. This supply, in
turn, powers two additional supplies, 5V and -12V.
6.2.4.3
All power supplies reside on the main PCB. There are three, as
explained in this section.
Power Supply, 12V -
supplies, the Velocimeter sensors and, through a microprocessor-
controlled power switch, the photocell probe.
The photocell power switch allows the microprocessor to turn
off the photocell when it is not needed, saving power and
extending battery life. The 12V supply also powers the analog
filters, azimuth conditioning, and contrast circuitry.
Power Supply, 5V -
and digital circuitry, including the microprocessor and memory.
Power Supply, -12V -
Velocimeters and analog filters, azimuth conditioning, and the
contrast circuitry. The -12V sensor power currently has a limit of
20 mA. It can take a continuous short without damage. The
effect of a short circuit on the microprocessor and display
depend on the battery voltage.
6.2.4.4
All supplies help operate the serial communications hardware.
Fuses protect the batteries and sensor power from short circuit
damage.
Honeywell Confidential: Use or disclosure of information on this page is subject to the restrictions on the title page.

On-Off Logic

Power Supplies
The 12V supply powers the other
The 5V supply runs the screen display
The -12V supply powers the
Short-circuit Protection