For The Vibrex 2000 (Basic) - Honeywell Vibrex 2000 User Manual

Digital dynamic balancer / analyzer systems
Table of Contents

Advertisement

5. Press "TRACK & BALANCE" from the main menu of
6. Acquire a balance measurement on channel 1A, and
7. Press Spectrum Analysis from the main menu.
8. Complete the following table by adjusting the EUT
6.3.5.5
Complete steps 4 through 8 and fill out the following table:
NOTE: For the 13590-2, there is no need to repeat the test
on channel 2.
Signal Gen
Frequency
40 Hz
5 Hz
Honeywell Confidential: Use or disclosure of information on this page is subject to the restrictions on the title page.
Adjust the amplitude of the sinewave on channel 1
so that 13.44 Vrms (+/- 0.01 Vrms) is shown on the
True RMS meter.
the EUT.
record the measurement parameters on the data sheet.
maximum frequency, the signal generator frequency and
the sinewave amplitude on the True RMS meters as
listed, and acquire a spectrum. Record the peak value in
the table.

For The Vibrex 2000 (Basic)

Vibrex
2000 Max
Freq
Balance
1,500
Chapter 6 – Maintenance
True RMS
Vibrex
Meter
2000 Peak
Amplitude
(IPS)
(MV)
Channel 1
13.44
13.44
13841 Rev A
221
Specification
(assumes 5%
for True
RMS meter)
0.90 <X<
1.11 11:30 -
12:30, 2400
RPM +/-3.6
0.90 <X<
1.11