Error List - Swann AMI Codes-II CC Operator's Manual

Monitoring system for the automatic, continuous measurement and dosing control of chlorine based on the dpd colorimetric method apha 4500 cl-g and on en iso 7393-2
Table of Contents

Advertisement

AMI Codes-II CC
Troubleshooting
7.3.
A-96.250.581 / 220616

Error List

Error
Non-fatal Error. Indicates an alarm if a programmed value is ex-
ceeded.
Such Errors are marked E0xx (bold and black).
Fatal Error
(blinking symbol)
Control of dosing devices is interrupted.
The indicated measured values are possibly incorrect.
Fatal Errors are divided in the following two categories:
 Errors which disappear if correct measuring conditions are re-
covered (i.e. Sample Flow low).
Such Errors are marked
 Errors which indicate a hardware failure of the instrument.
Such Errors are marked
HOLD
14:10:45
R1
0.22 ppm
0.26 ppm
R2
0.04 ppm
28 B/s
25.4°C
Main Menu
Messages
Diagnostics
Maintenance
Operation
Installation
Messages
Pending Errors
Maintenance List
Message List
Pending Errors
Error Code
Alarm low
<Enter> to Acknowledge
E0xx
(bold and orange)
E0xx (bold and
Error or
Error not yet acknowledged.
Check Pending Errors 1.1.5 * and
take corrective action.
Press [ENTER].
Reagent level low
1
Indicates the remaining reagent in per-
cent
Navigate to menu Messages.
Press [ENTER].
Navigate to menu Pending Errors.
1.1
Press [ENTER].
Press [ENTER] to acknowledge the
1.1.5
Pending Errors. The Error is reset and
E002
saved in the Message List.
* Menu numbers see
Program Overview, p. 73
red)
fatal Error
69