Table of Contents

Advertisement

AMI Codes-II
Troubleshooting
7.3.
66

Error List

Error
Non-fatal Error. Indicates an alarm if a programmed value is ex-
ceeded.
Such Errors are marked E0xx (bold and black).
Fatal Error
(blinking symbol)
Control of dosing devices is interrupted.
The indicated measured values are possibly incorrect.
Fatal Errors are divided in the following two categories:
 Errors which disappear if correct measuring conditions are re-
covered (i.e. Sample Flow low).
Such Errors are marked
 Errors which indicate a hardware failure of the instrument.
Such Errors are marked
HOLD
14:10:45
R1
0.15 ppm
R2
23 B/s
25.4 °C
Main Menu
Messages
Diagnostics
Maintenance
Operation
Installation
Messages
Pending Errors
Maintenance List
Message List
Pending Errors
Error Code
Alarm low
<Enter> to Acknowledge
E0xx
(bold and orange)
E0xx (bold and
Error or
Error not yet acknowledged.
Check Pending Errors 1.1.5 * and
take corrective action.
Press [ENTER].
Reagent level low
1
Indicates the remaining reagent in per-
cent.
To see the pending Errors:
Navigate to menu Messages.
Press [ENTER].
Navigate to menu Pending Errors.
1.1
Press [ENTER].
Press [ENTER] to acknowledge the
1.1.5
Pending Errors. The Error is reset and
E002
saved in the Message List.
* Menu numbers see
Program Overview, p. 70
red)
fatal Error
A-96.250.571 / 050517

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the AMI Codes-II and is the answer not in the manual?

Questions and answers

Nonhlanhla
March 2, 2025

Using the AMI Codes-II, there is Chlorine in the system, yet the system reads 0.00. Reseted, still no feedback. what could be the problem.

1 comments:
Mr. Anderson
March 2, 2025

If the Swann AMI Codes-II system reads 0.00 for chlorine despite chlorine being present, possible issues could include:

1. Reagent Issues – The required reagents (OXYCON ON-LINE DPD, buffer solutions, etc.) might be depleted, expired, or improperly mixed.
2. Clogged or Dirty Components – The sample inlet, inlet filter, or flow regulating valve might be clogged, preventing proper water flow.
3. Calibration or Sensor Issues – The system may need recalibration, or the sensor might be faulty or improperly installed.
4. Power or Electrical Issues – The instrument may not be receiving proper power, or there could be an electrical fault affecting measurement.
5. Interference from Additives – Certain water additives like corrosion inhibitors or cyanuric acid might interfere with chlorine detection.

Checking reagent levels, cleaning components, recalibrating, and verifying power and water flow can help identify and resolve the issue.

This answer is automatically generated

Nonhlanhla
March 2, 2025

Please share Data Sheet for the Pherispatic Pump

Table of Contents