AMU-II Oxytrace

Error List

7.
44
Error List
Error
Non-fatal error. Indicates an alarm if a programmed value is
exceeded.
Such Errors are marked E0xx (bold and black).
Fatal error
(blinking symbol)
Control of dosing devices is interrupted.
The indicated measured values are possibly incorrect.
Fatal errors are divided in the following two categories:
 Errors which disappear if correct measuring conditions are
recovered (i.e. Sample Flow low).
Such errors are marked
 Errors which indicate a hardware failure of the instrument.
Such errors are marked
HOLD
14:10:45
R1
7.04 ppm
R2
8 l/h
25.4°C
Messages
Pending Errors
Maintenance List
Message List
Pending Errors
Error Code
Alarm low
<Enter> to Acknowledge
E0xx (bold and
E0xx (bold and
Error or
Error not yet acknowledged.
Check Pending Errors 1.1.5 and take
corrective action.
Navigate to <Messages>/<Pending
1.1
Errors>.
Press [Enter] to acknowledge a pend-
1.1.5
ing errors.
E002
 The error is reset and saved in
the message list.
orange)
red)
fatal error
A-96.150.541 / 111223