Calibration; Prepare The Standard Solution; Start The Calibration - Swann AMI Phosphate-II Operator's Manual

Hide thumbs Also See for AMI Phosphate-II:
Table of Contents

Advertisement

AMI Phosphate-II
Maintenance
6.5.
6.5.1
6.5.2
42

Calibration

Prepare the Standard Solution

To prepare the standard solution proceed as follows:
1 Put a pipette in the standard solution 1000 ppm
2 Fill the pipette with 1 ml of standard solution.
3 Put the pipette into a volumetric flask and empty it.
4 Fill the volumetric flask with one liter demineralized water.

Start the calibration

Calibration
Close sample tap
- prepare 1 liter of standard.
Standard: 1.00 ppm
Wait for next prompt
Calibration
- Fill flow cell with
standard
<Enter> to continue
Calibration
Wait till calibration is
finished
Progress
Calibration
Calibration completed
Wait till flow cell is empty
Wait for next prompt
Calibration
Operation completed
Open sample tap
Slope
<Enter> to save
1 Navigate to menu Maintenance /
3.1.1
Calibration.
2 Press [Enter].
3 Follow the dialog on the Display.
3.1.1
3.1.1
3.1.1
Press [Enter], to save the calibration or
3.1.1
[Exit] to discard.
1.006
A-96.250.641 / 290121

Advertisement

Table of Contents
loading

Table of Contents