Table of Contents

Advertisement

AMU-II Oxytrace
Maintenance
6.4.
A-96.150.541 / 111223

Calibration

The sensing part of the sensor must not be in direct contact with
water!
In the wet flow cell, the atmosphere will be saturated with water
vapor. This atmosphere will produce the most accurate calibration
results.
The required time for a calibration depends mainly on the difference
between temperature and oxygen content in the sample and in the
air. It can take 15–20 minutes. This is also the case if the electrolyte
was exchanged.
As soon as the reading is stable, the microprocessor will store the
calibration data in the memory.The end of the calibration is indicated
on the display.
To perform a calibration proceed as follows:
1 Navigate to menu <Maintenance> /<Calibration>.
2 Press [Enter] to start the calibration and follow the dialog on the
display.
Calibration
Close regulating valve
to turn off sample flow.
<Enter> to continue
Calibration
Take sensor out of
flow cell and dry
membrane and sensor
<Enter> to continue
3 Stop the sample flow at with the flow
3.1.5
regulating valve.
4 Unscrew and remove the threaded
sleeve [A], see
change, p.
5 Remove the oxygen sensor [B] from
3.1.5
the flow cell [C].
6 Dry the sensor membrane and the
flow cell with a soft paper tissue.
Electrolyte ex-
32.
35

Advertisement

Table of Contents
loading

Table of Contents