Siemens sitrans LU01 Instruction Manual

Siemens sitrans LU01 Instruction Manual

Ultrasonic long-range level monitoring system
Table of Contents

Advertisement

Instruction Manual May 2008
sitrans
LU01

Advertisement

Table of Contents
loading

Summary of Contents for Siemens sitrans LU01

  • Page 1 Instruction Manual May 2008 sitrans LU01...
  • Page 2 The user is responsible for all changes and repairs made to the device by the user or the user’s agent. • All new components are to be provided by Siemens Milltronics Process Instruments Inc. • Restrict repair to faulty components only.
  • Page 3: Table Of Contents

    Table of Contents ..............................1 Introduction SITRANS LU01 ............................1 SITRANS LU01 Features ........................3 Programmable Features ......................3 The Manual ...............................4 ...............................5 Specifications Electronics ..............................5 Programmer ..............................6 Transducer ..............................7 Options ...............................7 Cable ................................7 Safety marking symbols ......................8 ..............................9 Installation SITRANS LU01 ............................9 Location ............................9 Cable/Conduit Entry Requirements ...................9...
  • Page 4 ...............................23 Operation Display in RUN Mode ...........................23 Keypad ..............................24 System Performance Evaluation .......................25 Performance Test Results ........................25 ..........................27 Parameter Reference Helpful Hints ..........................27 Parameter Reset Features ........................28 Programming Security .........................28 Quick Start (P001 to P007) ........................29 Application Parameters ........................32 Volume (P050 to P055) .........................32 Display and Reading (P060 to P062) ....................36 Failsafe (P070 to P072) ........................37 Relays (P100 to P119) ...........................39...
  • Page 5 Echo Processing Displays (Scope Displays, P810) .............86 Dolphin Plus Display ........................86 Distance Calculation ..........................86 Sound Velocity ............................86 Volume Calculation ..........................87 Universal, Linear (P050 = 9) .......................87 Universal, Curved (P050 = 10) ....................88 Maximum Process Speed ........................88 Application Examples ...........................89 Example 1 - Level Measurement .....................90 Material Volume ..........................92 Example 2 - Space Measurement ...................92 Application Assistance ......................94...
  • Page 7: Introduction

    Introduction SITRANS LU01 Note: • The SITRANS LU01 is to be used only in the manner outlined in this instruction manual. • This product is intended for use in industrial areas. Operation of this equipment in a residential area may cause interference to several frequency based communications.
  • Page 8 Programming can be done locally using the portable programmer keypad, or remotely through optional Dolphin Plus software or SmartLinx. • The programmer transmits the keypad entries via infrared link to the SITRANS LU01, and can be removed when not in use. •...
  • Page 9: Sitrans Lu01 Features

    SmartLinx provides protocol specific hardware and software for interface with popular industrial communication systems. The SITRANS LU01 is typically used to monitor material level in open or closed vessels but can be used in almost any process which requires a distance measurement (within the system range).
  • Page 10: The Manual

    0-20, 4-20, 20-0, or 20-4 mA with an adjustable range and over-range limits The Manual The manual is designed to help you get the most out of your SITRANS LU01, and it provides information on the following: • Product specifications •...
  • Page 11: Specifications

    • custom graphics backlit LCD with 51 mm (2 in) x 127 mm (5 in) viewing area Synchronization: • up to 16 SITRANS LU01 units can be synchronized together Temperature Compensation • –50 to 150 °C (–58 to 302 °F) •...
  • Page 12: Programmer

    • 67 mm W x 100 mm H x 25 mm D (2.6" W x 4" H x 1" D) • ABS plastic Weight • 150 g (0.3 lb) The use of approved watertight hubs/glands is required for Type 4X / NEMA 4X, IP65 on watertight applications. Page 6 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02...
  • Page 13: Transducer

    • maximum length 3 m (10 ft) Allen-Bradley is a registered trademark of Rockwell Automation. Modbus is a registered trademark of Schneider Electric. DeviceNet is a trademark of ODVA (Open DeviceNet Vendor Association). 7ML19985BE02 SITRANS LU01 – INSTRUCTION MANUAL Page 7...
  • Page 14: Safety Marking Symbols

    • refer to the associated instruction manual. Safety marking symbols In manual On Product Description (Label on product: yellow background.) Caution: refer to accompanying documents (manual) for details. Alternating Current Direct Current Earth (ground) Terminal Protective Conductor Terminal Page 8 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02...
  • Page 15: Installation

    The following procedure applies to all SITRANS LU01 level monitor installations. See Application Examples on page 89 for additional installation requirements. Also, refer to the instruction manuals of all other equipment connected to the SITRANS LU01 for additional installation instructions.
  • Page 16: Mounting

    Mounting Inspect all cartons and packaging for possible damage during shipment, before removing the SITRANS LU01 and associated equipment. Loosen the 6 enclosure lid (captivated) screws and swing the lid open. Remove the 4 Board B mounting screws (outer corners) and remove the circuit board assembly.
  • Page 17: Interconnection

    Connect all associated equipment cable shields to the SITRANS LU01 shield connections. To avoid differential ground potentials, do not connect cable shields to ground (earth) elsewhere. Insulate or tape cable shields at all shield junctions to prevent ground loops.
  • Page 18 •All field wiring must have insulation suitable for at least 250V. •Hazardous voltage present on transducer terminals during operation •Relay contact terminals are for use with equipment having no accessible live parts and wiring having insulation suitable for at least 250V. Page 12 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02...
  • Page 19: Optional Smartlinx Module

    Refer to the SmartLinx module documentation for any required hardware Note: settings prior to replacing the LCD card or closing the SITRANS LU01 lid. Wire in the SmartLinx card according to SmartLinx Manual. Replace the LCD card and secure in place using the screws removed in Step 1.
  • Page 20: System Diagram

    Transducer cables must be run in a grounded metal conduit separate from other wiring (except TS-3 temperature sensor wiring, if applicable). • Hazardous voltage present on transducer terminals during operation. to transducer Page 14 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02...
  • Page 21: Temperature Sensor

    21 22 23 24 25 not exceeding the rating of the relays. relays are Note: shown in de-energized n.o. n.c. state. to customer’s equipment See Specifications for ratings. mA Outputs OUTPUTS 0/4-20mA isolated output to Ω 7ML19985BE02 SITRANS LU01 – INSTRUCTION MANUAL Page 15...
  • Page 22: Communication

    Refer to the appropriate SmartLinx manual for installation and wiring. Level System Synchronization Avoid mounting the SITRANS LU01 near another ultrasonic level monitor. Likewise, when more than one monitor is installed within a single plant/facility, ensure the transducer cables of each system are run in separate grounded metal conduits. If this system separation is impractical, or if measurement difficulties are encountered, system synchronization may be required.
  • Page 23: Power

    • Never operate the SITRANS LU01 with the enclosure lid open, or with the ground (earth) wire disconnected. • Ensure that any associated alarm or control equipment is disconnected until satisfactory operation is verified.
  • Page 24: Programmer

    The hand programmer fits into the docking bay and is kept there with a magnet. Use the hand programmer to change individual parameters Communications Access Communications link is through the internal RJ-11 port RJ-11 port. RJ-11 cable Page 18 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02...
  • Page 25: Programming

    Number. The Index refers to the Relay Number as identified by the Index Type indicators. Parameter Numbers have a preset Parameter Value for each Index Number. Program the SITRANS LU01 to obtain the desired RUN mode operation. Display In PROGRAM mode, the Index Type, Index Number, Parameter Number, and Parameter Value (as well as a variety of other programming information) may be viewed.
  • Page 26: Keypad

    Keypad In PROGRAM mode, use the SITRANS LU01 programmer keys to perform the identified functions. numeric key values function keys Description DISPLAY: shift access to Index, Parameter Number, or Parameter Value dis- play. NUMBERS: enter the numeric value into the accessed display.
  • Page 27: Program Mode Entry

    Relay status and mA output values are held at last known values (unless affected by a parameter alteration or transducer firing) until RUN mode is re- entered. RUN mode is automatically re-entered if the SITRANS LU01 is left unattended in PROGRAM mode for an extended period. Changing Parameters Enter the new value, and press ENTER .
  • Page 28: Parameter Reset Features

    Index and Parameter Number displayed press CLEAR and ENTER To reset all parameters to preset values, use Master Reset (P999). Perform a Master Reset (P999) if the SITRANS LU01 was bench tested Note: using arbitrary Parameter Values before system installation, following an EPROM replacement, or whenever complete reprogramming is required.
  • Page 29: Operation

    Operation In RUN mode, the SITRANS LU01 detects material levels and provides control functions. The SITRANS LU01 automatically starts in RUN mode when power is applied. Display in RUN Mode In the RUN mode, the following values and indicators are observed. Many indicators are specific to certain operating conditions and so not all indicators are not displayed at any given time.
  • Page 30: Keypad

    Auxiliary Reading "distance" (material level to transducer face). initiates PROGRAM mode access (see switches the Reading between "Units/percent of Span" (completes PROGRAM mode access). stops/starts the Point Number auto display scroll. Page 24 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02...
  • Page 31: System Performance Evaluation

    0 until the next valid measurement is made. If the Failsafe Time Left reaches 0, the SITRANS LU01 flashes LOE in the Reading display. All associated data is supplied to the Peripheral Communications terminals (27 and 28).
  • Page 32 If the SITRANS LU01 provides accurate and repeatable measurements, however alternate Reading units, failsafe action, relay, or mA output operation is desired, proceed to Application Parameters on page 32. If all operating conditions cannot be observed during the System Performance Evaluation, refer to Enhancement Parameters Reading Measurement (P920).
  • Page 33: Parameter Reference

    Parameter Reference The SITRANS LU01 is configured through its parameters, and the application determines the parameter values which are entered into the unit. Please check your value entries carefully before operating the SITRANS LU01 to ensure optimum performance. Helpful Hints Please note the following: •...
  • Page 34: Parameter Reset Features

    To reset all parameters to preset values, refer to Master Reset (P999) on page 84. Note: bench tested Perform a Master Reset (P999) if the SITRANS LU01 was using arbitrary Parameter Values before system installation, following an EPROM replacement, or whenever complete reprogramming is required.
  • Page 35: Quick Start (P001 To P007)

    Transducer Scanning (P726 to P729) • P905 Transmit Pulse Use a setting just fast enough to keep up with your process. Slower settings provide higher accuracy. Faster settings allow for more level fluctuations. 7ML19985BE02 SITRANS LU01 – INSTRUCTION MANUAL Page 29...
  • Page 36 P004 Transducer Specifies the Siemens Milltronics transducer connected to the unit. No transducer attached ST-25 ST-50 ST-100 LR-21 LR-13 ST-H XCT-8 XPS-10 Values XCT-12 XPS-15 XPS-30 XPS-40 XLT-30 XLT-60 XLS-30 XLS-60 XRS-5 • P842 Short Shot Frequency • P843 Long Shot Frequency •...
  • Page 37 For distance operation (P001 = 3), Span is preset to the Empty value (P006). Enter a lower value if desired. If the automatic setting is not high enough, mount the transducer higher (see Installation/Transducer Mounting on page 10) and enter the new Empty (P006) distance. 7ML19985BE02 SITRANS LU01 – INSTRUCTION MANUAL Page 31...
  • Page 38: Application Parameters

    Application Parameters Volume (P050 to P055) Use these parameters to enable the SITRANS LU01 to show readings based on vessel volume (rather than level). P050 Tank Shape Enter the Tank Shape value matching the monitored vessel or wet well. When Operation is level (P001 = 1), liquid (material) volume is calculated. Alternatively, when Operation is space (P001 = 2), remaining vessel capacity is calculated.
  • Page 39 Tank Shape (P050) value. Make sure selected chosen units allow LCD volume display. Note: Examples: If max. volume = 3650 m , enter 3650 If max. volume = 267500 gallons, enter 267.5 (thousands of gallons) 7ML19985BE02 SITRANS LU01 – INSTRUCTION MANUAL Page 33...
  • Page 40 Breakpoint Primary Index Range: 0.0 to 9999 Values • P055 Volume Breakpoints Related Enter the following: • up to 32 level breakpoints (where volume is known) if P050 = 9 or 10 Page 34 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02...
  • Page 41 For more on Volume Characterization, go to page 87. P055 Volume Breakpoints and Characterization (Universal Volume Calculation) Each segment defined by the level breakpoints (P055) requires a volume so that the SITRANS LU01 can make the level-to-volume calculations. Breakpoint Primary Index Range: 0.0 to 9999 Values •...
  • Page 42: Display And Reading (P060 To P062)

    To calculate true volumes see Volume (P050 to P055). • Avoid entering a value that, when multiplied by the maximum current Reading, exceeds the display capabilities. If value exceeds four digits, EEEE is shown. Page 36 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02...
  • Page 43: Failsafe (P070 To P072)

    Failsafe (P070 to P072) As preset, in the event of a measurement or technical difficulty, the SITRANS LU01 holds the Reading, Bar Graph, mA outputs, and relays at their last known values. To operate process control equipment under these conditions, alter the following parameters as required.
  • Page 44 If the timer expires (before a valid measurement is made), the SITRANS LU01 advances to the Failsafe Material Level (P071) as restricted by Failsafe Advance (P072). When a valid measurement is made after the timer expires, the SITRANS LU01 advances to the new material level (if changed), as restricted by Failsafe Advance (P072) and the timer resets.
  • Page 45: Relays (P100 To P119)

    P701 Max Empty Rate Relays (P100 to P119) The SITRANS LU01 has four relays (or digital outputs) used to control devices and alarms. While the number of devices is limited by the relays, all control functions are accessible through software and each parameter is indexed to the relays.
  • Page 46 Set Up 1 Set Up 2 Values Set Up 3 Set Up 4 • P110 Level Source Alters • P001 Operation Related Programming the relays independently is the most common method used. Note: Page 40 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02...
  • Page 47 Notes: • OFF is displayed if the Relay Set Up (P100) selected does not use the Alarm parameter accessed. • Ch is displayed if a Custom Relay Parameter was previously operator altered. 7ML19985BE02 SITRANS LU01 – INSTRUCTION MANUAL Page 41...
  • Page 48: Custom Relays (P111 To P113)

    P111 Relay Control Function Sets the control algorithm used to trip the relay. When accessed, the Parameter Type display changes to the Relay symbol and the Index display changes to the Relay Number (corresponding to the SITRANS LU01 terminals). Relay Primary Index...
  • Page 49: Independent Relay Setpoints

    Celsius (°C). Rate alarms are entered in Units / minute or percent of Span per minute. A positive value is entered for a filling rate alarm; negative for emptying. 7ML19985BE02 SITRANS LU01 – INSTRUCTION MANUAL Page 43...
  • Page 50 Stop pump up on level at 80% * Values shown are for illustration purposes only. Enter values which apply to your particular installation. ! 2% factory set deadband, adjustable via P116 Page 44 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02...
  • Page 51 The distance above and below the bound alarm setpoints. Relay Primary Index Range: 0.000 to 9999 Values Preset: 2% of span • P111 Relay Control Function • P112 Relay ON Setpoint Related • P113 Relay OFF Setpoint 7ML19985BE02 SITRANS LU01 – INSTRUCTION MANUAL Page 45...
  • Page 52: Independent Relay Failsafe (P129)

    50 – all pump controls To select an independent Relay Failsafe value: Press FUNCTION to display the Auxiliary Function symbol. Press ARROWS to scroll through the failsafe options. Select option and press ENTER Page 46 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02...
  • Page 53: Ma Output (P200 To P219)

    Output (P200 to P219) When an mA Output Parameter is accessed, the mA symbol is displayed in the Index Type field and the mA output number (corresponding to the SITRANS LU01 terminals) is displayed in the Index field. P200 mA Output Range Determines the mA output range.
  • Page 54: Independent Ma Setpoints (P210 And P211)

    Sets the minimum mA output value (in mA) to be produced. mA output Primary Index Range: 0.000 to 22.00 Values Preset: 0.0 or 3.8 • P200 mA Output Range Related • P213 mA Output Max Limit Page 48 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02...
  • Page 55: Ma Output Trim (P214 To P215)

    Calibrates the 20 mA output. mA output Primary Index Range: -1.00 to 1.000 Values Preset: 0.000 • P214 4 mA Output Trim Related Adjust this value so the device indicates 20.00 mA when P215 is accessed. 7ML19985BE02 SITRANS LU01 – INSTRUCTION MANUAL Page 49...
  • Page 56: Ma Output Failsafe (P219)

    Shows the highest temperature encountered, as measured by the temperature sensor in the transducer (if applicable). Range: - 50 to 150°C (view only) Values Preset: - 50°C Press CLEAR keys to reset the log after a short circuit on the transducer wiring. Page 50 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02...
  • Page 57: Profile Records (P330 To P337)

    These parameters are for authorized service personnel or WARNING: technicians familiar with Siemens Milltronics echo processing techniques. These features can record up to ten Echo profiles, initiated manually (P330), or automatically (P331 et al) for viewing at a later time using Dolphin Plus or an oscilloscope.
  • Page 58: Auto Record On And Off Setpoints (P334 To P337)

    If ---- is displayed for either P334 or P335, Auto Profile Records are saved regardless of current level (subject to all other restrictions). Enter the level value in Units (P005) or percent of Span (P007) as referenced to Empty (P006). Page 52 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02...
  • Page 59 P336 Auto Record Filling / Emptying If the LOE condition exceeds the period entered, the Echo Profile is saved. When set for 0 LOE is not required for an Auto Profile Record to be saved. 7ML19985BE02 SITRANS LU01 – INSTRUCTION MANUAL Page 53...
  • Page 60: Installation Records (P340 To P342)

    Values • P341 RUN Time Related • P342 Start Ups P341 RUN Time View the number of days this SITRANS LU01 has been in operation. Range: 0.000 to 9999 (view only) Values • P340 Date of Manufacture Related • P342 Start Ups The RUN Time value is updated once a day, and cannot be reset.
  • Page 61 In gasses other than air, the temperature variation may not correspond Note: with the speed of sound variation. Turn off temperature sensor and use a fixed temperature. 7ML19985BE02 SITRANS LU01 – INSTRUCTION MANUAL Page 55...
  • Page 62 The units used are m/s if P005 = 1, 2, or 3, or ft/s if P005 = 4 or 5. Page 56 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02...
  • Page 63: Temperature Compensation (P660 To P664)

    • P661 Temp Fixed The SITRANS LU01 uses the TS-3 temperature sensor assigned to the transducer. If one is not connected, the ultrasonic/temperature transducer is used. If the transducer does not have an internal temperature sensor, the Temp Fixed (P661) value is used.
  • Page 64: Rate (P700 To P707)

    Rate (P700 to P707) These parameters determine how material level changes are reported. P700 Max Fill Rate Adjusts the SITRANS LU01 response to increases in the actual material level (or advance to a higher Failsafe Material Level, P071). Range: 0.000 to 9999 Values •...
  • Page 65 This is automatically altered along with Maximum Process Speed (P003). This value automatically alters the Rate Update Time (P705) and / or Rate Update Distance (P706). Alternatively, these parameter values may be altered independently. 7ML19985BE02 SITRANS LU01 – INSTRUCTION MANUAL Page 59...
  • Page 66: Measurement Verification (P710 To P713)

    • P713 Echo Lock Window This value (in % of Span, P007) is automatically altered when Maximum Process Speed (P003) is altered. The higher the value entered, the greater the fluctuation stabilized. Page 60 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02...
  • Page 67 When set for max verification or material agitator, a new measurement outside of the Echo Lock Window (P713) must meet the sampling criterion (P712). For total lock, Echo Lock Window (P713) is preset to zero 0. The SITRANS LU01 continuously searches for the best echo according to the algorithm chosen (P820). If the selected echo is within the window, the window is then centered about the echo.
  • Page 68: Transducer Scanning (P725 To P729)

    Sync terminal. P727 Scan Delay The delay, in seconds, between measurements from transducer points . Range: 0.000 to 60.00 Values Preset: 5.0 • P003 Maximum Process Speed Altered by • P001 Operation Related Page 62 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02...
  • Page 69: Display (P730 To P733)

    If necessary, enter the Parameter Number to default in the Auxiliary Reading display. That value will show in the auxiliary reading area by default. Other values are available but will reset to the parameter defined here. 7ML19985BE02 SITRANS LU01 – INSTRUCTION MANUAL Page 63...
  • Page 70 / untag any accessed parameter. is displayed to indicate the parameter accessed is tagged. Quick Start parameters (P001 – P007) and those changed from factory Note: default settings cannot be untagged. Page 64 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02...
  • Page 71: Peripheral Communication Support Parameters (P740 To P749)

    Values P749 Serial Bus Type Displays the type of communication bus the SITRANS LU01 is currently set for. If there is no connection, the SITRANS LU01 defaults to RS-485. If a bus type other than RS-485 is connected, the SITRANS LU01 displays the type of bus it is, or is attempting to, communicate with.
  • Page 72: Communications (P772)

    Communications (P772) The SITRANS LU01 communication ports are configured by a series of parameters that are indexed by port. Communication parameters are indexed to these communication ports, unless otherwise noted: Port Description RS-232 port (RJ-11 modular telephone) RS 485 port on terminal block P772 Baud Rate The communication rate with the master device.
  • Page 73: Echo Processing (P800 To P807)

    P792 Bus Error Count A count that increments by 1 each time a bus error (P791) is reported.. Range: 0* to 9999 Values Error count; provide this number to your Siemens Milltronics repre- sentative for troubleshooting. • P790 Hardware Error...
  • Page 74 When a transducer with a submergence shield is submerged, the shield traps an air pocket that creates a special echo. The SITRANS LU01 recognizes the echo and advances the reading to the highest level and operates displays and outputs accordingly.
  • Page 75 Increases SITRANS LU01 response when the monitored surface is close to the transducer face. Select short and long to have short and long acoustic shots fired for each measurement, regardless of the transducer to surface distance. Select short to have only short shots fired if the Echo Confidence (P805) produced by a short shot exceeds the short Confidence Threshold (P804) and the monitored surface is always within the Short Shot Range (P852).
  • Page 76: Advanced Echo Processing (P810 To P825)

    See in the Troubleshooting Section on page 99. Advanced Echo Processing (P810 to P825) The following parameters are for authorized Siemens Milltronics Service Note: personnel or technicians familiar with Siemens Milltronics echo processing techniques. Anatomy of an Echo Profile The relevant parts of an echo profile are listed here.
  • Page 77 • Echo Lock Window display OFF Use the Scope Displays after pressing TRANSDUCER to observe the result of parameter alterations. Take several measurements to verify repeatability and overcome Echo Lock (P711) restrictions. 7ML19985BE02 SITRANS LU01 – INSTRUCTION MANUAL Page 71...
  • Page 78: Profile Pointer (P817 To P825)

    • P817 Profile Pointer Time • P819 Profile Pointer Amplitude • P820 Algorithm • P821 Spike Filter Related • P822 Narrow Echo Filter • P823 Reform Echo • P825 Echo Marker Trigger Page 72 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02...
  • Page 79 Press MEASURE to display the Auxiliary Function symbol. Press ARROWS to access the desired Reading display symbols. Press ENTER when the required algorithm is displayed. 7ML19985BE02 SITRANS LU01 – INSTRUCTION MANUAL Page 73...
  • Page 80 P817 Profile Pointer Time • P818 Profile Pointer Distance • P819 Profile Pointer Amplitude Related • P820 Algorithm • P821 Spike Filter • P822 Narrow Echo Filter • P825 Echo Marker Trigger Page 74 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02...
  • Page 81: Advanced Tvt Adjustment (P830 To P835)

    Echo Profile at the sharpest rising portion of the Echo Profile representing the true echo. This value is preset to 90% when P002=1 (Liquid) or 50% when P002=2 (Solid). Advanced TVT Adjustment (P830 to P835) The following parameters are for authorized Siemens Milltronics Service Note: personnel or technicians familiar with Siemens Milltronics echo processing techniques.
  • Page 82 DISPLAY to access the Breakpoints. Press MODE to select the Auxiliary function. Press to scroll the breakpoint index through points 1 to 40, and move the curve pointer right or left respectively. Page 76 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02...
  • Page 83 Echo Profile. Alternatively, if TVT Type is set for TVT Slopes (P830 = 6), preset is 2000. Use this parameter to adjust the slope declination when TVT type is set to TVT slopes (P830=6). 7ML19985BE02 SITRANS LU01 – INSTRUCTION MANUAL Page 77...
  • Page 84: Advanced Shot Adjustment (P840 To P852)

    Advanced Shot Adjustment (P840 to P852) These parameters are for Siemens Milltronics service personnel only. Note: P840 Short Shot Number The number of short shots to be fired (and results averaged) per transmit pulse. Range: 0 to 100 Values Preset: 1 •...
  • Page 85 • P803 Shot / Pulse Mode • P840 Short Shot Number • P842 Short Shot Frequency Related • P844 Short Shot Width • P851 Short Shot Floor • P852 Short Shot Range 7ML19985BE02 SITRANS LU01 – INSTRUCTION MANUAL Page 79...
  • Page 86: Test (P900 To P913)

    • P851 Short Shot Floor This feature is automatically altered when Transducer (P004) is altered. Test (P900 to P913) These parameters are for Siemens Milltronics service personnel only. Note: P900 Software Revision Number View the EPROM Rev. #. Range: 00.00 to 99.99 (view only)
  • Page 87 CPU into an infinite loop to test the watchdog timer. On successful completion (10 seconds) the RUN mode is entered and the SITRANS LU01 is reset. Programming is kept and the unit responds as if there had been a power failure.
  • Page 88 Err is displayed if the transducer is not equipped with an internal temperature sensor. P913 Sensor Temperature Access this parameter to display the temperature in °C (as monitored by the TS-3). Range: -50 to 150 Values OPEn is displayed if a TS-3 is not connected. Page 82 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02...
  • Page 89: Measurement (P920 To P927)

    The distance in Units (P005) or % of Span (P007) between Empty (P006) and the monitored surface. P922 Space Measurement The distance between the monitored surface and Span (P007). P923 Distance Measurement The distance between the monitored surface and the transducer face. 7ML19985BE02 SITRANS LU01 – INSTRUCTION MANUAL Page 83...
  • Page 90 . C.ALL displays until the reset is complete. CAUTION: Be careful when using this feature. All data will be reset. For convenience, be sure to record the values you want to re-enter. Page 84 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02...
  • Page 91: Technical Reference

    Technical Reference Transmit Pulse The SITRANS LU01 transmit pulse consists of one or more electrical shot pulses, which are supplied to the scanning relay. The scanning relay is activated as required, to supply the transmit pulse to the Transducer(s) connected to the SITRANS LU01 terminals.
  • Page 92: Echo Processing Displays (Scope Displays, P810)

    The sound velocity of the transmission medium is affected by the type, temperature, and vapour pressure of the gas or vapour present. As preset, the SITRANS LU01 assumes the vessel atmosphere is air at 20°C (68°F). Unless altered, the sound velocity used for the distance calculation is 344.1 m/s (1129 ft/s).
  • Page 93: Volume Calculation

    Velocity (P653) may be entered directly. Volume Calculation The SITRANS LU01 provides a variety of volume calculation features (P050 to P055). If the vessel to be monitored does not match any of the 8 preset Tank Shape calculations, a Universal Volume calculation may be used. Use the level/volume graph or chart provided by the vessel fabricator (or create one based on the vessel dimensions).
  • Page 94: Universal, Curved (P050 = 10)

    (as well as 1 breakpoint exactly at the angle) on the curve. Maximum Process Speed The SITRANS LU01 Maximum Process Speed (P003) to material level changes is designed to exceed the most demanding installation requirements.
  • Page 95: Application Examples

    The Process Speed setting automatically adjusts various parameters affecting the SITRANS LU01 response to material level changes as in the table that follows. Values Dependent on Measurement Response Parameter (P003) Value Name (Units) 1(slow) 2(medium) 3(fast) 4(surge) (5imm.) P070 Failsafe Timer (min)
  • Page 96: Example 1 - Level Measurement

    Example 1 - Level Measurement Material Level This is the most common application of the SITRANS LU01 level monitor. For this example we'll assume the following: • one 30 m high cement silo is to be monitored. • the maximum vessel filling rate is 0.08 m per minute.
  • Page 97 = 10.00 m. 12:27 am High alarm on, actual level = 21.36 m, reported level = 25.00 m. 12:47 am Infeed stopped, actual level = 22.96 m, reported level = 27.00 m. 7ML19985BE02 SITRANS LU01 – INSTRUCTION MANUAL Page 91...
  • Page 98: Material Volume

    XCT-8 transducer is mounted in a standpipe so that the face is 18 in. above the vessel top. • a TS-3 temperature sensor is mounted in the vessel to monitor liquid temperature. 3.00ft (P007) 3.00ft (P007) 1.5ft 12.00 ft (P052) (P053) Page 92 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02...
  • Page 99 Note that on power off the mA output drops to 0.00 mA. During a power interruption, the last known mA output value is maintained in memory for 1 hour minimum. If power is interrupted beyond 1 hour, on power resumption, the mA output immediately assumes the new value. 7ML19985BE02 SITRANS LU01 – INSTRUCTION MANUAL Page 93...
  • Page 100: Application Assistance

    Full in feet) set the mA Function (P201) to 2. Application Assistance The preceding examples describe only a few ways in which the SITRANS LU01 can be applied to process measurement requirements. The SITRANS LU01 can be used to monitor almost any process (within the temperature, measurement range, and chemical immunity capabilities of the system) where a distance measurement or determination of presence vs.
  • Page 101: Mt-00 Measurement Message

    26 through 32. The mode of communication is established by the configuration of the connection made at the SITRANS LU01, either RS-232, RS-485 or bipolar current loop. simplex convention The SITRANS LU01 uses communication protocol. Data messages are continuously transmitted at regular intervals (no poll is required) at 4800 baud. All data is transmitted in structured messages of ASCII characters consisting of 8 data bits, no parity, and 1 stop bit.
  • Page 102: Mt-01 Hold Message

    Field Name Definition Start Of Message STX ($02) 2 characters, 01 (indicates the SITRANS LU01 was removed from the Message Type RUN mode) End Of Message CR ($0D) Page 96 SITRANS LU01 – INSTRUCTION MANUAL...
  • Page 103: Mt-03 Point Not Scanned

    Start Of Message STX ($02) Message Type 2 characters, 03 (indicates the Point Number is not in service) Point # 3 characters, 001 to 003 (e.g. Point Number 1) End Of Message CR ($0D) 7ML19985BE02 SITRANS LU01 – INSTRUCTION MANUAL Page 97...
  • Page 104: Maintenance

    Maintenance The SITRANS LU01 should require no maintenance or cleaning, though good housekeeping practices in and around the area of the enclosure are recommended. Wipe out the area of the enclosure lid docking bay recess with a clean dry cloth (if necessary) before installing the programmer.
  • Page 105: Troubleshooting Guide

    Reading is erratic, with True echo too weak or Relocate and/or re-aim transducer at little or no relation to wrong echo being pro- material level or object material level. cessed. Proceed to Measurement Difficulties. 7ML19985BE02 SITRANS LU01 – INSTRUCTION MANUAL Page 99...
  • Page 106: Measurement Difficulties

    If a measurement difficulty occurs for greater than the Failsafe Timer (P070) setting, LOE is flashed alternately with the Reading display. Under certain conditions, a measurement difficulty may cause the SITRANS LU01 to lock on to a false echo and report a fixed or wrong reading.
  • Page 107: Fixed Reading

    Sometimes, material simply cannot be detected during vessel filling. In these cases, set up failsafe operation so the SITRANS LU01 anticipates the rate of material level increase and adjusts the reported Reading accordingly. This operation is reliable, as long as when the dust (or foam) settles a valid echo can be received.
  • Page 108: Wrong Reading

    SITRANS LU01 shield terminals and not to ground elsewhere. If the SITRANS LU01 is mounted in close proximity to (or transducer cables run near those of) another ultrasonic level monitor, see Level System Sync (P726).
  • Page 109 (P852) to stabilize the shot mode used for the echo evaluation. Also, adjust the Short Shot Bias to increase (or decrease) the amount of preference given to short shot echoes over long shot echoes. Should a stable measurement still not be attainable, contact Siemens Milltronics or your local distributor. 7ML19985BE02 SITRANS LU01 –...
  • Page 110: Programming Charts

    P072 Failsafe Level Advance Relays P100 Relay Set Up (G) P101 Hi Alarm P102 Lo Alarm P103 Hi Hi Alarm P104 Lo Lo Alarm P111 Relay Function P112 Relay A Setpoint P113 Relay B Setpoint Page 104 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02...
  • Page 111 P340 Date of Manufacture P341 Run Time P342 Start Ups Range Calibration P650 Offset Calibration P651 Sound Velocity Calibration P652 Offset Correction P653 Velocity P654 Velocity @ 20° C Temperature Compensation P660 Temp Source 7ML19985BE02 SITRANS LU01 – INSTRUCTION MANUAL Page 105...
  • Page 112 P740 Peripheral Communications (G) P748 RS485 Termination P749 Serial Bus Type SmartLinx P772 Baud Rate P790 Hardware Error P791 Bus Error P792 Bus Error Count Echo Processing P800 Near Blanking P801 Range Extension Page 106 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02...
  • Page 113 P841 Long Shot Number P842 Short Shot Frequency P843 Long Shot Frequency P844 Short Shot Width P845 Long Shot Width P850 Short Shot Bias P851 Short Shot Floor P852 Short Shot Range 7ML19985BE02 SITRANS LU01 – INSTRUCTION MANUAL Page 107...
  • Page 114 Siemens Milltronics Process Instruments Inc. Siemens Milltronics Process Instruments Inc. 2008 1954Technology Drive, P .O. Box 4225 Subject to change without prior notice Peterborough, ON, Canada K9J 7B1 Rev. 2.4 Tel: (705) 745-2431 Fax: (705) 741-0466 *7ml19985BE02* Email: techpubs.smpi@siemens.com...