Perform a One-Port Reflection Calibration
A one-port reflection calibration uses known standards to correct for
directivity, source match, and frequency response errors when using
narrowband detection. When you perform a one-port reflection
calibration, the analyzer performs correction at each data point across
the selected frequency band. The default number of data points per
sweep is 201, but you can select any number between 3 and 1601.
Interpolation recalculates the error correction array for reduced
frequency spans. If the frequency span is increased, the calibration is
invalidated, and the default one-port calibration is automatically
restored.
To perform a one-port reflection calibration, you will need one of the
following calibration kits depending on the nominal impedance of your
analyzer:
Cal Kit Model
Number
HP 85039B
HP 85036B
HP 85036E
HP 85032B
HP 85032E
HP 85031B
HP 85033D
HP 85038A
HP 85038F
HP 85038M
ET User's Guide
Connector
Impedance
Type
75 Ω
type-F
75 Ω
type-N
75 Ω
type-N
50 Ω
type-N
50 Ω
type-N
50 Ω
APC-7
50 Ω
3.5 mm
50 Ω
50 Ω
50 Ω
Measuring Reflection Response
For Use
with Male
Test Ports
N/A
7-16
7-16
7-16
Making Measurements
For Use
with Female
Test Ports
N/A
3-33