Figure 3-12
NOTE
Making Measurements
Measuring Transmission Response
22. When the instrument prompts you to connect a through cable, insert
the female-to-female adapter between the test cables connected to the
REFLECTION port and the TRANSMISSION port as shown in
Figure
3-12. Press
Through-Cable Connection
23. The analyzer will measure each standard and then calculate new
calibration coefficients. The message "Calibration complete."
will appear for a few seconds when the analyzer is done calculating
the new error correction array.
24. Remove the adapter between the test cables connected to the
REFLECTION port and the TRANSMISSION port.
Changing sweep frequencies (and other source parameters) may affect
your calibration. See
3-30
Measure Standard
Chapter 6
for more information.
.