NOTE
Getting Started
Performing the Operator's Check
Make a Reflection Measurement
1. Leave the cable connected to the analyzer as shown in
The quality of the cable will affect these measurements; make sure you
use a cable with the characteristics described in
page
2-10.
2. Press
PRESET
Reflection
SCALE
3. Press
POWER
4. Press
Default 1-Port
CAL
5. Verify that the data trace falls completely below −16 dB. See
Figure 2-9
for a typical HP 8714ET result. The HP 8712ET should
look similar, but end at 1300 MHz.
2-14
User Preset
or
Factory Preset
Enter
.
10
0
dBm
.
.
Figure
2-6.
"Equipment List" on
MEAS 1
ET User's Guide