HP 8712ET User Manual page 250

Rf network analyzers
Hide thumbs Also See for HP 8712ET:
Table of Contents

Advertisement

NOTE
NOTE
Calibrating for Increased Measurement Accuracy
Calibration Kits
280
!
LOSS
290
!
300
END
How to download a calibration kit ASCII file. After creating a
calibration kit ASCII file, you can download the program by following
these steps:
Calibration kit definitions must be in DOS format. LIF format is not
supported for cal kit definitions.
1. Insert the disk with the cal standard definitions into the analyzer's
built-in disk drive.
2. Press
SAVE RECALL
the screen. If it is not "INT:\" (for internal disk), press
Internal 3.5" Disk Prior Menu
3. Highlight the file (CALKIT, in the case of the Example Programs
Disk).
4. Press
Recall State
Loaded cal kit information for 4 calibration standards
Recall of cal kit from CALKIT complete
5. Press
More Cal Cal Kit
CAL
front panel knob or the
port in the display's table. Next, select the softkey that corresponds to
the User Cal Kit where the ASCII program stored the calibration kit
(default is User Cal Kit A).
Cal kit coefficients are displayed in the cal kit block of the operating
parameters screen; press
Operating Parameters
6-24
2.2E+9 $ Ohms/Sec
and note the first word of the second line on
.
and wait for the prompt:
More Cal Kits
keys to highlight the desired
SYSTEM OPTIONS
press
Next Screen
;
Select Disk
, and use the
a total of four times.
ET User's Guide

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 8714et

Table of Contents